Fluke Calibration 5322A Operator's Manual page 130

Multifunction electrical tester calibrator
Hide thumbs Also See for 5322A:
Table of Contents

Advertisement

5322A
Operators Manual
[SOUR]:SAF:IDD[:CURR]:INST?
Description: This command returns the instantaneous measured current
Query:
[SOUR]:SAF:IDD[:CURR]:RES?
Description: This command returns the measured current flowing through the
Query:
[SOUR]:SAF:RCDT:TIME(?) <DNPD>
Description: This command sets the nominal value of trip time. If not already
Parameters: <DNPD>
Example:
Query:
[SOUR]:SAF:RCDT:CURR?
Description: Returns the measured trip current.
Query:
122
flowing through the Product. The Product returns the current value
using exponential format. Example: 1 mA is returned as 1.0000e-
03. If not already selected, this command switches the Product
into the Leakage Current Differential function mode.
SAF:IDD:INST?
DUT. Product returns the current value using exponential format.
If not already selected, this command switches the Product into
the Leakage Current Differential function mode.
SAF:IDD:RES?
selected, this command switches the Product into the RCD Trip
Time function mode.
SAF:RCDT:TIME 0.055
SAF:RCDT:TIME?
Product returns measured trip current value using exponential
format. Example: 0,25A is returned as 250.00e-03. If not already
selected, this command switches the Product into the RCD Trip
Time function mode.
SAF:RCDT:CURR?
1mA is returned as 1.0000e-03
Parameter represents the value of
the leakage current expressed in
seconds. See the Specifications
(found online at www.Flukecal.com)
for acceptable ranges.
RCD Trip Time mode 55ms.
The Product returns the nominal
value of trip time using exponential
format. Example: 20 ms is returned
as 20e-03.