Fluke Calibration 5322A Operator's Manual page 15

Multifunction electrical tester calibrator
Hide thumbs Also See for 5322A:
Table of Contents

Advertisement

Loop, Line, and Ground Bond Resistance
The Product sources high power rated low resistance standards from 14 mΩ to
1.7 kΩ, which are especially suited for loop impedance, line impedance, and
ground bond resistance testing. Ground bond outputs can be configured as either
2-wire or 4-wire. The Product senses and displays Device Under Test (DUT) test
conditions, types of test current, and current levels up to 40 A. In the 5322A/VLC,
an Active Loop Compensation module will compensate for residual resistance
when performing loop and line impedance calibrations.
Residual Current Device (RCD) Testing
In the RCD function, the Product acts as a circuit breaker to calibrate trip time in
the range of 10 ms to 5 s, and trip current in the range of 3 mA to 3 A. All tested
parameters of the DUT are scanned and displayed on the Product display.
Leakage Current Source
The Product sources simulated leakage current from 0.1 mA to 30 mA with a
maximum compliance voltage up to 250 V ac. Leakage current modes include
touch, substitute, and differential current.
AC/DC Voltage Generation (5322A/VLC and 5322A/5/VLC Only)
When equipped with the ac/dc voltage calibrator, the Product is capable of
calibrating the voltmeter function on many electrical safety testers. The output
voltage range is 3 V to 600 V both ac and dc. The ac frequency range is 40 Hz to
400 Hz. This voltage source also generates stable power line voltage for
powering appliance testers.
Meter Functions
The Product is equipped with a built-in low frequency voltmeter and ammeter.
The voltmeter measures up to 5000 V ac rms or V dc, while the ammeter
measures up to 30 A. The Meter also measures ac power, including phase, and
dc power.
HIPOT Function
In the HIPOT sub function, the Product can measure total harmonic distortion of
ac signals up to 5000 V ac rms and ripple coefficient of dc signals up to
5000 V dc. The function is suited for HIPOT tester parameter verification. The
Product also has the capability to verify HIPOT timers up to 999 s.
Multifunction Electrical Tester Calibrator