Select The Line Voltage; Install The Product In An Equipment Rack; Switch Power On - Fluke Calibration 5322A Operator's Manual

Multifunction electrical tester calibrator
Hide thumbs Also See for 5322A:
Table of Contents

Advertisement

Select the Line Voltage

The Product operates on two different input line voltages. The line voltage setting
is indicated on the face of the line voltage switch found on the Product rear
panel. See Rear-Panel Features.
To change the line voltage:
1. Disconnect the Product from line power by unplugging the line power cord.
2. Using a flat-blade screwdriver, rotate the switch until the desired voltage is
under the arrowhead on the line voltage switch.
3. Ensure the proper Line Power fuse for the selected line voltage selection is
installed in the Product. See Line Power Fuses.
4. Attach the Product to the power source using a line power cord appropriate
for the mains outlet.

Install the Product in an Equipment Rack

The Product can be mounted in a standard-width, 24-inch (61-cm) deep
equipment rack. To mount the Product in an equipment rack, use the Rack
Mount Kit, Model Y5320A, instructions are included with the kit.

Switch Power On

With the Product setup for the proper line voltage, push the power switch on the
back panel to the I side of the switch. See Rear-Panel Features. During its
power-up cycle, the Product shows a power supply test screen while initializing
internal circuits and checking the mains connection. The mains connection tests
are:
• Power line voltage test – The line voltage must be within preset limits. For the
230 V setting, the range must be between 180 V and 260 V. For the 115 V
setting, the limits are between 90 V to 130 V.
The Product requires standard nonsymmetrical power line mains
(NT) with line (hot) wire, protection earth, and neutral wire.
• Power line frequency test – The frequency must be within preset limits: 49 Hz
to 51 Hz or 59 Hz to 61 Hz.
• Potential difference and polarity test – The potential difference between
neutral and protective earth must be less than 15 V.
Multifunction Electrical Tester Calibrator
Note
Install the Product in an Equipment Rack
13