Set the Calibration Password

The calibration default password is set to 2235. Only when the password is
changed to a non-zero value must a password be entered to access the
calibration mode. The purpose of the password is to prevent unauthorized users
from changing the calibration settings.
To set the calibration code:
1. After you enter the setup menu, push  or  or turn the rotary knob to
highlight Calibration Password in the list of setup selections.
2. Push the Select softkey or push in the rotary knob.
3. If the calibration password is set to a non-zero value, you will be prompted to
enter the present password.
4. Use the keyboard to enter a new code and push . Make sure what is
shown in the display and what you think you have entered are the same.
After completing the next step, you must have the correct password to gain
access to the calibration functions.
5. Push the Write softkey to set the calibration password and return to the setup
menu. To exit setting the password without changing it, push Exit.

Set the Time

To set the time:
1. After you enter the setup menu, push  or  or turn the knob to highlight
Time in the list of setup selections.
2. Push the Select softkey or push in the rotary knob.
3. Use  or  to position the cursor on the least significant digit of the hours,
minutes, or seconds. The cursor only rests on the least significant digit.
4. Use  or  or turn the rotary knob to change the time element.
5. Push the Exit softkey to accept the date and exit the date setting function.
Multifunction Electrical Tester Calibrator
Calibrator Setup Menu
29