Calibrator
Calibration Manual
July 2012
© 2011, 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke Calibration 5080A

  • Page 1 5080A Calibrator Calibration Manual July 2012 © 2011, 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Page 3: Table Of Contents

    Introduction ......................1 Safety Information ..................... 1 Instruction Manuals ................... 3 5080A Operators Manual ................3 5080A Getting Started Manual ..............4 5080A Calibration Manual ................4 Accessories and Options ..................4 How to Contact Fluke Calibration ..............5 General Maintenance ..................5 How to Replace the Line Fuse ...............
  • Page 4 5080A Calibration Manual Calibration Steps ................... 54 How to Calibrate Through the Remote Interface ..........57 Remote Commands ..................58 Generating a Calibration Report ..............63 Replaceable Parts ....................63 How to Obtain Parts ..................63 Parts ....................... 64...
  • Page 5 Leveled Sine Wave Function Verification: Flatness ..........27 Leveled Sine Wave Function Verification: Frequency .......... 30 Marker Generator Function Verification ..............30 Test Equipment Required for 5080A Calibration ..........31 Test Equipment Required for DC Volts Calibration ..........32 Calibration Steps for DC Volts ................33 Test Equipment Required for AC Volts Calibration ..........
  • Page 6 5080A Calibration Manual Calibration Steps for DC Current ................35 Test Equipment Required for AC Current Calibration .......... 38 Calibration Steps for AC Current ................38 Calibration Steps for AUX DC Current ..............40 Calibration Steps for AUX AC Volts ..............41 Test Equipment Required for Resistance Calibration ..........
  • Page 7 List of Figures Figure Title Page Fuse Access ......................6 Current Fuse Compartment ..................8 Accessing the Air Filter..................10 Low Resistance Verifications Connections ............18 High Resistance Source Calibration Connections ..........19 High Resistance Verification for 1 GΩ and higher ..........20 DC Voltage Adjustment Setup ................
  • Page 8 5080A Calibration Manual...
  • Page 9: Introduction

    Introduction  Warning If the 5080A Calibrator is operated in a manner not specified by this manual or other documentation provided by Fluke, the protection provided by the Calibrator may be impaired. The 5080A Calibrator is a fully programmable precision source of the following: •...
  • Page 10 5080A Calibration Manual Symbols used on the Calibrator and in this manual are explained in Table 1. Table 1. Symbols Symbol Description Symbol Description   AC (Alternating Current) Earth Ground   Important Information: refer to manual Shock Hazard Complies with relevant North ...
  • Page 11: Instruction Manuals

    To avoid damage to the Calibrator, do not apply voltage in excess of the marked rating to any terminal. Instruction Manuals The 5080A Manual Set gives complete data for operators and service or maintenance technicians. The set includes: • 5080A Operators Manual (provided on CD-ROM) •...
  • Page 12: 5080A Getting Started Manual

    5080A Calibration Manual 5080A Getting Started Manual The 5080A Getting Started Manual contains a brief introduction to the 5080A Manual Set, instructions on how to get your Calibrator prepared for operation and a complete set of specifications. 5080A Calibration Manual This 5080A Calibration Manual includes calibration and verification procedures to keep the Calibrator within specifications.
  • Page 13: How To Contact Fluke Calibration

    Manual Calibration Software. MET/BASE and MET/TRACK required. Manual MET/CAL Options can be ordered factory installed with a new calibrator (5080A/MEG, 5080A/SC, and 5080A/SC/MEG), or added later at a Fluke service center for an additional installation and calibration charge. How to Contact Fluke Calibration To contact Fluke Calibration, call one of the telephone numbers shown below.
  • Page 14 5080A Calibration Manual 2. Put a screwdriver blade in the tab at the left side of fuse holder and pry until it can be removed with the fingers. 3. Remove the fuse from the compartment for replacement or verification. Be sure the correct fuse is installed.
  • Page 15: How To Replace The Current Fuses

    Calibrator General Maintenance How to Replace the Current Fuses The two Calibrator current outputs have fuse protection. If the Calibrator cannot source current, one or both of the current fuses can be blown.  Warning To prevent electric shock, turn the Calibrator off and remove the mains power cord.
  • Page 16 5080A Calibration Manual gew368.eps Figure 2. Current Fuse Compartment 4. Remove and examine the fuses as necessary. Table 4 shows the part number and rating of each fuse. Table 4. Current Fuses Current Output Fuse Description Fluke Part Number 4A/500V Ultra-Fast Blow (F3)
  • Page 17: How To Clean The Air Filter

    General Maintenance How to Clean the Air Filter  Warning To prevent risk of injury, do not operate or power the 5080A calibrator without the fan filter in place.  Caution Damage can occur when the Calibrator becomes too hot if the area around the fan is restricted, the intake air is too warm, or the filter becomes clogged.
  • Page 18: How To Clean Outside Surfaces

    5080A Calibration Manual oq062f.eps Figure 3. Accessing the Air Filter How to Clean Outside Surfaces To clean the outside surfaces, wipe the case, front-panel keys, and lens using a soft cloth slightly dampened with water or a non-abrasive mild cleaning solution that does not harm plastics.
  • Page 19: Performance Tests

    Fluke PM6680B How to Verify the Main Calibrator Use Tables 6 through 16 to make sure the 5080A operates to its specifications. The tables are for approved metrology personnel who have access to a standards laboratory that has equipment that can test calibration equipment of this level of accuracy. The tables show the recommended test points and the upper and lower limits for each point.
  • Page 20 5080A Calibration Manual Table 6. Verification Tests for DC Voltage (Normal) Range Output Lower Limit Upper Limit 329.999 mV 0.000 mV -0.0100 mV 0.0100 mV 329.999 mV 329.000 mV 328.9538 mV 329.0462 mV 329.999 mV -329.000 mV -329.0462 mV -328.9538 mV 3.29999 V...
  • Page 21 Calibrator Performance Tests Table 8. Verification Tests for DC Current Range Output Lower Limit Upper Limit 330 μA 0.00 μA -0.100 μA 0.100 μA 330 μA 329.00 μA 328.670 μA 329.330 μA 330 μA -329.00 μA -329.330 μA -328.670 μA 3.3 mA 0.00000 mA -0.000250 mA...
  • Page 22 5080A Calibration Manual Table 9. Verification Tests for 4-Wire Resistance Range Output Tolerance 0.0000 Ω 0.0000 Ω ±0.01000 Ω 1.0000 Ω 1.0000 Ω ±0.00989 Ω 1.9000 Ω 1.9000 Ω ±0.00931 Ω 10.000 Ω 10.000 Ω ±0.014 Ω 19.000 Ω 19.000 Ω...
  • Page 23 Calibrator Performance Tests Table 10. Verification Tests for AC Voltage (Normal) (cont.) Range Output Frequency Lower Limit Upper Limit 3.3 V 3.2900 V 1 kHz 3.28653 V 3.29347 V 33 V 32.900 V 45 Hz 32.8686V 32.9314V 33 V 32.900 V 100 Hz 32.8620 V 32.9380 V...
  • Page 24 5080A Calibration Manual Table 12. Verification Tests for AC Current (cont.) Range Output Frequency Lower Limit Upper Limit 3.3 mA 3.2900 mA 45 Hz 3.28219 mA 3.29781 mA 3.3 mA 3.2900 mA 100 Hz 3.28186 mA 3.29814 mA 3.3 mA 3.2900 mA...
  • Page 25: Megohm Option Verification Tests

    Calibrator Performance Tests Table 14. Verification Tests for Phase (Normal Voltage vs. Aux Voltage) Range, AUX Output Lower Upper Range Output Frequency Output Phase Limit Limit Voltage 3.3 V 329 mV 65 Hz 3.29999 V 329 mV 0 ° -0.25 ° 0.25 °...
  • Page 26 5080A Calibration Manual Fluke 5080A Fluke 8508A gru001.eps Figure 4. Low Resistance Verifications Connections 2. Push  on the Calibrator. 3. Push the softkey labeled MODE until lvr shows above the right-most Calibrator softkey. 4. Set the multimeter for true ohms and auto range.
  • Page 27: High Resistance Source (Hvr) Verification

    To verify the High Resistance Source function: 1. Connect the high resistance source output to the input terminals of a standard multimeter in 2-wire configuration as shown in Figure 5. Fluke 5080A Fluke 8508A gru002.eps Figure 5. High Resistance Source Calibration Connections 2.
  • Page 28 Calibrator and connect the LO terminal on the megohmmeter to the HI terminal on the calibrator. Turn the ground on and make the measurement. Fluke 5080A Megohmmeter gru003.eps Figure 6. High Resistance Verification for 1 GΩ and higher...
  • Page 29: Scope Option Verification Tables

    ±0.55 GΩ Scope Option Verification Tables Before the 5080A/SC Option leaves the Fluke factory, it is verified to meet its specifications at the test points shown in Tables 20 through 30. The verification test points are provided here as a guide when re-verification is desired. Table 44 is a list of test equipment necessary for SC200 Scope option calibration.
  • Page 30 5080A Calibration Manual Table 20. Voltage Function Verification: AC Voltage into a 1 MΩ Load (cont.) Measured Value 1-Year Spec. Nominal Value (p-p) Frequency Deviation (mV) (p-p) (mV) 20.0 mV 10 kHz 0.15 50.0 mV 10 kHz 0.23 89.0 mV 10 Hz 0.32...
  • Page 31 Calibrator Performance Tests Table 21. Voltage Function Verification: AC Voltage into a 50 Ω Load (cont.) Measured Value 1-Year Spec. Nominal Value (p-p) Frequency Deviation (mV) (p-p) (mV) 10.0 mV 100 Hz 0.12 10.0 mV 1 kHz 0.12 10.0 mV 10 kHz 0.12 20.0 mV...
  • Page 32 5080A Calibration Manual Table 22. Voltage Function Verification: DC Voltage into a 50 Ω Load (cont.) Nominal Value (dc) Measured Value (dc) Deviation (mV) 1-Year Spec. (mV) -25.0 mV 0.16 55.0 mV 0.24 -55.0 mV 0.24 100.0 mV 0.35 -100.0 mV 0.35...
  • Page 33 Calibrator Performance Tests Table . Voltage Function Verification: DC Voltage into a 1 MΩ Load (cont.) Nominal Value (dc) Measured Value (dc) Deviation (mV) 1-Year Spec. (mV) -450.0 mV 1.22 500.0 mV 1.35 -500.0 mV 1.35 3.3 V 8.35 -3.3 V 8.35 4.0 V 10.10...
  • Page 34 5080A Calibration Manual Table 25. Wave Generator Function Verification: 1 MΩ Load (cont.) Nominal Measured Deviation Waveform Frequency 1-Year Spec. Value (p-p) Value (p-p) (mV) Sine 6.5 V 10 kHz 195.10 mV Sine 55.0 V 10 kHz 1.65 V Triangle 5.0 mV...
  • Page 35 Calibrator Performance Tests Table 27. Leveled Sine Wave Function Verification: Amplitude Nominal Value 1-Year Spec. Frequency Measured Value (p-p) Deviation (mV) (p-p) (mV) 5.0 mV 50 kHz 0.300 10.0 mV 50 kHz 0.400 20.0 mV 50 kHz 0.600 40.0 mV 50 kHz 1.000 50.0 mV...
  • Page 36 5080A Calibration Manual Table 28. Leveled Sine Wave Function Verification: Flatness (cont.) Nominal Value 1-Year Spec. Frequency Measured Value (p-p) Deviation (mV) (p-p) (mV) 10.0 mV 10 MHz 0.25 10.0 mV 20 MHz 0.25 10.0 mV 50 MHz 0.25 10.0 mV 100 MHz 0.25...
  • Page 37 Calibrator Performance Tests Table 28. Leveled Sine Wave Function Verification: Flatness (cont.) Nominal Value 1-Year Spec. Frequency Measured Value (p-p) Deviation (mV) (p-p) (mV) 400.0 mV 1 MHz 6.10 400.0 mV 1 MHz 6.10 400.0 mV 2 MHz 6.10 400.0 mV 5 MHz 6.10 400.0 mV...
  • Page 38 5080A Calibration Manual Table 28. Leveled Sine Wave Function Verification: Flatness (cont.) Nominal Value 1-Year Spec. Frequency Measured Value (p-p) Deviation (mV) (p-p) (mV) 5.5 V 125 MHz 110.00 5.5 V 160 MHz 110.00 5.5 V 200 MHz 110.00 Table 29. Leveled Sine Wave Function Verification: Frequency Nominal Value 1-Year Spec.
  • Page 39: Mainframe Calibration

    Note Mixed in with the "output and measure" procedures are internal 5080A calibration procedures that are done without operator aid. 4. The Calibrator computes a software correction factor and stores it in volatile memory.
  • Page 40: Start Calibration

    It must be set to ENABLE to store the correction factors into nonvolatile memory. After you push the 5080A CAL softkey, the procedure works as follows: 1. The Calibrator automatically programs the outputs and prompts you to make external connections to applicable measurement instruments.
  • Page 41: Ac Volts Calibration (Normal Output)

    5. Measure and type in the values into the UUT for steps 1 through 9 in Table 33 as prompted. You Disconnect and reconnect the DMM as prompted during these steps. DC Voltage adjust completed. Table 33. Calibration Steps for DC Volts Step 5080A Output (NORMAL) 1.000000 V 3.000000 V -1.000000 V -3.000000 V 0.0000 mV...
  • Page 42 AC Measurement Standard To calibrate the ac voltage function: 1. Measure the 5080A output through Input 1 of a Fluke 5790A AC Measurement Standard. Use a Type N to dual banana adapter as shown in Figure 8. 2. Type in the measured values into the Calibrator for each step in Table 35 as prompted.
  • Page 43: Dc Current Calibration

    9. Exit calibration and save the calibration constants changed so far through the front panel menus or the CAL_STORE remote command. Table 37. Calibration Steps for DC Current Step 5080A Output (AUX, HI, LO) Shunt to Use 300.000 μA Fluke 742A-1k 1 kΩ Resistance Standard 3.00000 mA...
  • Page 44: Ac Current Calibration

    DC characterization can be done with the Calibrator, if you do the full 5080A dc current calibration first. When you do a dc characterization, data is obtained for each of the ac current levels needed by the ac current calibration procedure.
  • Page 45 5. Put the 5080A in OPERATE and measure the ac voltage across the shunt. 6. Calculate the ac current from the data derived from the dc characterization and the ac correction factors supplied by the shunt manufacturer.
  • Page 46 Current Shunt, 2 A Fluke A40A-20A Current Shunt, 20 A Fluke 792-7004 A40 Current Shunt, Adapter Table 39. Calibration Steps for AC Current 5080A Output (AUX HI, LO) Steps Amplitude Frequency Shunt to Use 3.2999 mA 55.00 Hz Fluke A40 10 mA 0.3300 mA...
  • Page 47 Calibrator Mainframe Calibration Fluke Corporation - Worldwide Support Center MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Fluke 5080A ACI ADJ DATE: 03-Oct-2011 AUTHOR: Gary Bennett, Metrology Specialist REVISION: ADJUSTMENT THRESHOLD: NUMBER OF TESTS: NUMBER OF LINES: CONFIGURATION: Fluke 5790A ============================================================================= STEP RANGE NOMINAL...
  • Page 48: Dc Volts Calibration (Aux Output)

    AUX HI and LO terminals on the UUT. Table 40 shows the calibration steps for AUX dc volts. Table 40. Calibration Steps for AUX DC Current Steps 5080A Output (Aux) 300.00 mV 3.0000 V 7.000 V...
  • Page 49: Ac Volts Calibration (Aux Output)

    AUX HI and LO terminals instead on the UUT. Table 41 shows the calibration steps for AUX ac volts. Table 41. Calibration Steps for AUX AC Volts 5080A Output (AUX) Steps Amplitude Frequency 300.00 mV...
  • Page 50 5080A Calibration Manual Table 43. Calibration Steps for Resistance Step 5080A Output (4-Wire Ohms, NORMAL and AUX) 1.0000 Ω 1.9000 Ω 10.000 Ω 19.000 Ω 100.000 Ω 190.000 Ω 1.00000 kΩ 1.90000 kΩ 10.0000 kΩ 19.0000 kΩ 100.000 kΩ 190.000 kΩ...
  • Page 51: Sc200 Calibration

    Figure 13. Connections for Calibrating 2-Wire Resistance SC200 Calibration The SC200 is an oscilloscope calibration option for the 5080A. The sections that follow explain how to calibrate the SC200 Oscilloscope Option. Note Although this procedure will calibrate the SC200 option in the field, it is recommended the calibrator be sent to Fluke for calibration and verification.
  • Page 52 5080A Calibration Manual Table 44. Required Test Equipment for SC200 Calibration (cont.) Test Equipment Recommended Model Minimum Specifications Edge Rise Time and Aberrations Verification High Frequency Tektronix 11801 with SC- Frequency 2 GHz Digital Storage 22/26 sampling head, or Resolution 4.5 mV to 2.75 V...
  • Page 53 Calibrator SC200 Calibration Table 44. Required Test Equipment for SC200 Calibration (cont.) Test Equipment Recommended Model Minimum Specifications Edge Frequency, AC Voltage Frequency Verification PM 6680 with PM 9678 20 ms to 150 ns, 10 Hz to 10 MHz: Frequency Counter option <1.6 ppm uncertainty BNC Cable...
  • Page 54: Calibration Steps

    5080A Calibration Manual Calibration Steps Note The Calibrator mainframe must be calibrated before you start to calibrate the SC200 option. Adjustments can be necessary after repair of the SC200. Hardware adjustments must be done before calibration. You must calibrate the option after hardware adjustments. See the Hardware Adjustment section.
  • Page 55 Calibrator SC200 Calibration Table 45. AC Square Wave Voltage and Edge Settings for HP3458A HP 3458A Settings Voltage Input Frequency NPLC Delay (topline) DELAY (baseline) 10 Hz 0.02 s 0.07 s 100 Hz 0.002 s 0.007 s 1 kHz 0.01 0.0002 s 0.0007 s 5 kHz...
  • Page 56 5080A Calibration Manual Note The Calibrator shows a warning message if the typed in measurement is out of limit. Look at the measurement again and re-enter the value carefully, with the correct multiplier (m, k, n, p, etc.). If the warning continues to show in the display, Calibrator repair can be necessary.
  • Page 57 Calibrator SC200 Calibration Note The topline is near 0 V, and the baseline is a negative voltage in the EDGE function. 4. At each calibration step, read measurements for a minimum of two seconds, with the HP 3458A MATH function to get the average or mean value. The “true amplitude”...
  • Page 58 5080A Calibration Manual Note The Calibration mainframe will warn when the typed in value is out of limits. If this warning occurs, inspect the setup and carefully type in the measurement again with the correct multiplier (m, u, n, p, etc.). If the warning continue to show in the display, repair can be necessary.
  • Page 59: Sc200 Hardware Adjustments

    Calibrator SC200 Calibration 15. Push the OPTIONS, then the STORE CONSTS softkeys to keep the new calibration constants. SC200 Hardware Adjustments You must make hardware adjustments to the leveled sine and edge functions each time the SC200 is repaired. This section contains a list of equipment and recommended models necessary for these procedures.
  • Page 60 5080A Calibration Manual -34 dBc -39 dBc 3rd harmonic 2nd harmonic yg127f.eps Figure 16. Leveled Sine Wave Harmonics Adjustment 8. If the second and third harmonics are not at the correct dB level, adjust R8 until they are. You can find that you can have the second harmonic at -34 dBc but the third harmonic is less than -39 dBc.
  • Page 61 Calibrator SC200 Calibration 1st Aberration 2nd Aberration 3rd Aberration om050f.eps Figure 17. Edge Aberrations Adjustment 8. Adjust R13 to set the edge signal at the 10 ns point to the reference level. 9. Adjust R12 for a flat edge signal. If necessary, adjust R13 again to keep the edge signal at the reference level.
  • Page 62: Megohm Option Calibration

    5080A Calibration Manual MEGOHM Option Calibration The sections that follow explain how to calibrate the MEGOHM Option. Note Although this procedure will calibrate the SC200 option in the field, it is recommended the calibrator be sent to Fluke for calibration and verification.
  • Page 63 Figure 18. MEGOHM Low Resistance Calibration Connections 2. Select the 4-wire TRUE ohms function on the multimeter. 3. Set the cal enable switch on the rear of the 5080A to the enable position. 4. Push , then Cal, Cal, Option Cal, and Megohm Cal.
  • Page 64 5080A Calibration Manual Table 49. MEGOHM Calibration Adjustment Points (cont.) Step MEGOHM Output 1.620 kΩ 6.810 kΩ 11.00 kΩ 20.00 kΩ 40.00 kΩ 77.00 kΩ 143.0 kΩ 274.5 kΩ 510.0 kΩ 980.0 kΩ 1.825 MΩ 3.490 MΩ 6.650 MΩ 9.760 MΩ...
  • Page 65: How To Calibrate Through The Remote Interface

    Calibrator How to Calibrate Through the Remote Interface Megohmmeter Fluke 5080A gru111.eps Figure 19. MEGOHM High Resistance Calibration Connections 8. Set the parameters of the QuadTech 1865 to the values shown in Table 50 and then do the adjustment steps in Table 51.
  • Page 66: Remote Commands

    These calibration commands can be sent through the IEEE-488 or serial interface. To type in commands through the serial interface: 1. Use a Fluke PM8914 cable to connect the applicable COM port on a PC to the 5080A Serial 1 connector.
  • Page 67 Calibrator How to Calibrate Through the Remote Interface stream are sequential commands. All commands that is not overlapped or coupled are sequential. Overlapped Commands – Commands that take added time to execute are overlapped commands. The execution time of these commands can overlap the subsequent command.
  • Page 68 Description Skip to next section of the calibration procedure. Example CAL_SECT CAL_START Description Start a calibration procedure. Parameters MAIN Procedure for the 5080A minus the scope cal option. ZERO Internal procedure to touch up zero offsets. OHMSZERO Internal procedure to touch up resistance offsets. SCOPE Procedure for the 5080A-SC200 scope cal option.
  • Page 69 Calibrator How to Calibrate Through the Remote Interface CAL_STATE? Description Gets the state of calibration. Response A calibration step is in progress. Waiting for a CAL_NEXT with reference (measurement) value. Instruction available, waiting for a CAL_NEXT. Not in a calibration procedure (or at end of one) CAL_STEP? Description Gets the name of the calibration step in progress.
  • Page 70 5080A Calibration Manual RPT? Description Gets a self-calibration report. Parameters 1. Type of report to print: STORED, ACTIVE, or CONSTS 2. Format of report: PRINT (designed to be read) SPREAD (designed to be loaded into a spreadsheet) 3. Calibration interval to be used for instrument specifications in the...
  • Page 71: Generating A Calibration Report

    “consts,” which is a listing of the active set of raw calibration constant values. Replaceable Parts This section contains an illustrated list of replaceable parts for the Fluke 5080A Calibrator. Parts are in the list by assembly and alphabetized by reference designator.
  • Page 72: Parts

    5080A Calibration Manual Parts Table 53 shows the replaceable parts for the 5080A Calibrator. Figure 20 shows the parts within the Calibrator. Table 53. Replaceable Parts Ref. Part Description Desig. Number 5080A-7603, MOTHER BOARD, A3 - TESTED 4018029 5080A-7605, OHMS, A5 - TESTED...
  • Page 73 (PAINTED) 5080A-2010,FRONT TRIM FIXING STRIP 3435237 5080A-8006,HANDLE, 4U 3468705 5080A-8001,KEYPAD RUBBER 5080A 3391346 5080A-4002,PCA, KEYPAD / ENCODER A1 / A2 3439674 CONTRACT MFG ITEM, JACK BAN SAFE BLK 2079863 CONTRACT MFG ITEM, JACK BAN SAFE RED 2079874 MP10 5080A-2019,HIGH VOLTAGE TERMINAL BLOCK...
  • Page 74 INSULATOR,NYLON,MOLDED,5/16 (M8),0.093 IN. MP24 3533983 LONG,BULK CABLE TIE,50LB,14.2 IN. LIGHT GREEN HIGH-TEMP NYLON MP25 3534000 WITH S-S LOCKING DEVICE,BULK MP26 5080A-8007,MAINS SWITCH PUSH ROD 5080A 3525824 MP27 5700A-2046 ,POWER BUTTON, ON/OFF 775338 MP28 5500A-8011 ,AIR FILTER 945287 MP29 5500A-2012 ,HOUSING, AIR FILTER...
  • Page 75 Number GUIDE,GETTING STARTED GUIDE, 5080A MP44 3502941 SHIPPING BOX,REGULAR SLOTTED CARTON,KRAFT,48LB MP45 3753014 B-C FLUTE,28.375,25.00,13.25 MP46 DVD,GNU LESSER GENERAL PUBLIC LICENSE - 5080A 3779181 LABEL,SERVICE ONLY LABEL- 5080A MP47 3779509 MP48 5080A-2020,SHIELD, A9 3669601 MP49 5440A-8198-01 ,BINDING POST, STUD, PLATED...
  • Page 76 5080A Calibration Manual gru105.eps Figure 20. Final Assembly...
  • Page 77 Calibrator Replaceable Parts gru106.eps Figure 20. Final Assembly (cont.)
  • Page 78 5080A Calibration Manual gru107.eps Figure 20. Final Assembly (cont.)
  • Page 79 Calibrator Replaceable Parts MP36, MP37 H121-H132 (x8) MP44 MP38, MP41 H132-H168 (x36) MP42 H132-H168 MP41 (x2) (x36) J1-J6 gru108.eps Figure 20. Final Assembly (cont.)
  • Page 80 5080A Calibration Manual...