Table of contents Introduction..................3 Section 1: Meter functions ............5 Direct voltage Alternating voltage Resistance Capacitance Direct current Alternating current Section 2: Temperature functions ..........12 Temperature Section 3: Power functions ............15 Power Power harmonics Section 4: Overload protection ..........20 Output Protection Safety warning: The 55XX Series calibrators are capable of producing...
Introduction Output Window displays main output Arrow Keys select highlighted digits in information. the Output Window. EDIT FIELD toggles between two values if present in the Separate STBY and OPR Keys control Output Window. output. Edit Knob, which has a detent for each SCOPE and Trig Out Keys enable the digit increment, allows you to vary the optional oscilloscope calibration func-...
Page 4
Each is eleven calibrators in one Compared with more traditional calibrators designed to calibrate a certain type of instrument, the Fluke Calibration 55XX Series rep- resents a whole new class of multi-product calibrator, offering an unprecedented range of dc and low-frequency electrical calibration workload coverage.
Section 1: Meter functions What you’ll need In addition to a 55XX Series Multi-Product Calibrator, you’ll need a Fluke 179 Digital Multimeter or equivalent and a set of compatible test leads. Direct voltage The 55XX Series calibrators can output direct voltage from 0 to ± 1020 V. 1.
Then turn the edit knob clockwise until the output display reads 33 V. Notice that the calibrator dropped into standby mode. This safety precaution protects the operator from inadvertently slewing the output to a potentially hazardous level. 7. Now press the key, located in the keypad to the left of the edit knob.
Page 7
3. Press the under WAVE sine. Notice the menu changes to WAVE tri for triangle wave. Also notice the output display has changed from a conventional rms voltage reading to peak-to-peak. 4. Press the under WAVE tri and the display changes to WAVE square.
4. Notice that the control window displays a range of lead com- pensation choices. Simply press the under COMP off to cycle through the choices—off, 2-wire and 4-wire. With compensa- tion off, the 55XX Series does not compensate for external lead resistances.
Page 9
output. The design of the 55XX Series calibrators’ synthesized capaci- tance circuit permits it to be used with meters that use both the dc charge/discharge method currently used on handheld dmms, as well as the ac technique used by RCL meters. 1.
Direct current 5502A CALIBRATOR PREV STBY EARTH SCOPE TRIG MENU EDIT SETUP RESET FIELD MEAS ZERO MULT • ENTER SHIFT POWER Direct current is available from 0 to ± 20.5 A. Sourcing direct current is identical to sourcing direct voltage; how- ever you have to change connections on the 55XX Series, and most likely also on the dmm.
Alternating current 5502A CALIBRATOR PREV STBY EARTH SCOPE TRIG MENU EDIT SETUP RESET FIELD MEAS ZERO MULT • ENTER SHIFT POWER Alternating current is available from 29 µA to 20.5 A over frequencies from 0.01 Hz to 30 kHz. Using a 55XX Series calibrator in alternating current mode is simi- lar to using it in alternating voltage mode;...
Section 2: Temperature functions What you’ll need In addition to a 55XX Series Multi-Product Calibrator, you’ll need a Fluke 54II or equivalent thermocouple thermometer, a K-type thermocouple and a K-type thermocouple wire with male connectors on each end.
Temperature 5502A CALIBRATOR PREV STBY EARTH SCOPE TRIG MENU EDIT SETUP RESET FIELD MEAS ZERO MULT • ENTER SHIFT POWER The 55XX Series calibrators have three temperature modes: • They can simulate eleven types of thermocouples (B C E J K L N R S T U) for calibrating thermometers and other thermocouple indi- cators.
Page 14
5502A CALIBRATOR THERMOMETER PREV STBY EARTH SCOPE TRIG MENU EDIT SETUP RESET CANCEL FIELD ˚C ˚F K HOLD T1-T2 CLEAR MEAS ZERO MEMORY PC / IR SEND LOGGING RECALL MULT • ENTER SHIFT POWER SETUP ENTER To simulate a thermocouple (source temperature) connect a thermo- couple cable (a piece of thermocouple wire with a male connector at each end) to the input of the 54 II and TC output of the 55XX Series calibrator.
Section 3: Power functions What you’ll need In addition to a 55XX Series Multi-Product Calibrator, you’ll need a Fluke 43B and a set of test leads. If you do not have access to a Fluke 43B, a general purpose watt meter with voltage and current inputs may be substituted.
Power POWER QUALITY ANALYZER 5502A CALIBRATOR PREV STBY EARTH SCOPE TRIG MENU ENTER EDIT SETUP RESET FIELD MEAS ZERO MENU RECORD MULT • ENTER SHIFT POWER HOLD SAVE PRINT Direct power can be simulated into phantom loads from 0 to 20.9 kW. This is accomplished by sourcing voltage from one output and cur- rent from another.
Page 17
3. Entering voltage plus voltage is the same as entering voltage plus current, except that you substitute voltage for current. The voltage output on the AUX terminal is limited to 7 V (1.0 mV to 5 V in ac mode). If you were simulating a current clamp that outputs 1 mV per amp, this output would simulate 100 W (10 V + 10 A) ac at 60 Hz: 4.
Page 18
6. Notice that in this mode, the 55XX Series calibrator’s control window displays a WAVE MENUS. When you press the under this display, WAVE MENU provides a range of choices, including harmonics, current and voltage waveforms and phase. First, we’ll demonstrate the waveform and phase control functions.
Power harmonics POWER QUALITY ANALYZER 5502A CALIBRATOR PREV STBY EARTH SCOPE TRIG MENU ENTER EDIT SETUP RESET FIELD MEAS ZERO MENU RECORD MULT • SHIFT ENTER POWER HOLD SAVE PRINT To cover harmonics analyzers like the Fluke 43B, the 55XX Series multi-product calibrators can source harmonics of the fundamental output frequency, up to the 50th harmonic.
Section 4: Overload Protection What you’ll need In addition to a 55XX Series Multi-Product Calibrator, you’ll need a Fluke 1587 Insulation Multimeter or equivalent.
Output protection 1587 INSULATION MULTIMETER 5502A CALIBRATOR Test Voltage HOLD RANGE MINMAX INSULATION TEST PREV STBY EARTH SCOPE TRIG MENU 50 V 1000 V EDIT SETUP RESET FIELD INSULATION MEAS ZERO 600V CAT INSULATION 1000V CAT MULT • ENTER SHIFT 440mA FUSED POWER...