5322A
Operators Manual
[SOUR]:SAF:LOOP:PONS(?) <CPD> { ON | OFF }
Description: This command sets the Power ON (Start Up) switch. This
Parameters: <CPD>
Example:
Query:
[SOUR]:SAF:LOOP:SRES(?) <DNPD>
Description: This command sets the series resistance of connection cables.
Parameters:
Example:
Query:
[SOUR]:SAF:LOOP:PENR(?) <CPD> { ON | OFF }
Description: This command sets the PE-N resistance switch. This command
Parameters: <CPD>
Example:
Query:
[SOUR]:SAF:LOOP:RESC
Description: Starts the internal measurement of power line residual impedance.
Example:
138
command switches the Product into the Loop Impedance function
mode. The state of the Power ON switch is the same for Loop and
Line functions.
SAF:LOOP:PONS ON
SAF:LOOP:PONS?
Parameter sets the value of series resistance in Ω.
<DNPD>
See the Specifications (found online at
www.Flukecal.com) for the acceptable ranges. This
command switches the Product into the Loop
Impedance function mode.
SAF:LOOP:SRES 0.07
SAF:LOOP:SRES?
switches the Product into the Loop Impedance function mode.
Parameter represents the switch state.
ON
OFF
SAF:LOOP:PENR ON
SAF:LOOP:PENR?
This command switches the Product into the Line Impedance
function mode.
SAF:LOOP:RESC
Parameter represents the switch state.
ON
Sets the power ON
OFF
Sets the power OFF
The Product returns set switch state.
The Product returns the value of series
resistance using exponential format.
Example: 0.72Ω is returned as 0.720e+00.
Sets the PE-N resistance switch ON
Sets the PE-N resistance switch OFF
The Product returns set switch state.
Need help?
Do you have a question about the 5322A and is the answer not in the manual?