5322A
Operators Manual
Terminators
To signify the end of a response sent to the controller, the Product sends a
terminator. For response message terminators, the Product sends the ASCII
character Line Feed with the EOI control line held high. The subsequent
characters are recognized by the Product as terminators when encountered in
incoming data:
• ASCII LF and CR characters
• Any ASCII character sent with the EOI control line true
Each command must end in <cr> or <lf>. Both codes <crlf> can be
used at the same time. The Product executes all commands written
on one line of the program after it receives <cr>, <lf> or <crlf> code.
Without this code, the program line is ignored.
Description of Abbreviations
<DNPD>
<CPD>
<SPD>
?
(?)
<cr>
<lf>
88
Note (only for USB)
Decimal Numeric Program Data - This format expresses decimal
number with or without the exponent.
Character Program Data - Usually, it represents a group of
alternative character parameters. For example {ON | OFF | 0 | 1}.
String Program Data - String value consisting from more parts. It
is used for date/time setting.
A flag that indicates a request for the value of the parameter
specified by the command. Do not use any other parameter than
the question mark.
A flag indicating a request for the parameter specified by the
command. This command permits a value to be set as well as
requested.
carriage return - ASCII code 13. This code executes the program
line.
line feed - ASCII code 10. This code executes the program line.