NOTE
Table 3-2
Model
16495F
16495G
16495H
16495J
16495K
Installation
Installing Accessories

To Connect Connector Plate

The available connector plates are listed in Table 3-2. Option 001 provides the
through connectors except for the Interlock connector that provides the soldering
patterns at the back side. For the option 002, the back of each connector is designed
for soldering. For connector plate installation information, refer to Agilent 16495
Installation Guide.
For the 16495H/J, the insulator is inserted between the panel and the outer
conductor of Triax connectors/BNC connectors.
Use the cables shown in Table 3-3 to connect from the B1500A connectors to the
connector plate's relative connectors. For the cable connections from the connector
plate to the DUT interface such as manipulator and probe card, see "Mounting
Connectors" on page 3-27 as reference.
Making non-Kelvin connection
The Force terminals can be used to force and measure DC voltage or current. If you
want to simplify the cable connections, open the Sense terminals and connect the
Force terminals only to the connector plate by using the triaxial cables.
If you make the Kelvin connection, use both Force and Sense terminals. Connecting
the Force and Sense lines together at the terminal of the device under test minimizes
the measurement error caused by the residual resistance of the connection cables.
The Kelvin connection is effective for the low resistance measurement and the high
current measurement.
Connector Plates
Option Item
Triax. (f-f) 12, GNDU (f-f) 1, Interlock (f) 1
16495F-001
Triax. (f) 12, GNDU (f) 1, Interlock (f) 1
16495F-002
Triax. (f-f) 24, GNDU (f-f) 1, Interlock (f) 1
16495G-001
Triax. (f) 24, GNDU (f) 1, Interlock (f) 1
16495G-002
Triax. (f-f) 6, Coax. (f-f) 6, GNDU (f-f) 1, Interlock (f) 1
16495H-001
Triax. (f) 6, Coax. (f) 6, GNDU (f) 1, Interlock (f) 1
16495H-002
Triax. (f-f) 8, Coax. (f-f) 4, GNDU (f-f) 1, Interlock (f) 1
16495J-001
Triax. (f) 8, Coax. (f) 4, GNDU (f) 1, Interlock (f) 1
16495J-002
16495K-001
Connector plate, rubber holder