Measurement accuracy
Measurement accuracy at the measurement port of the MFCMU or the MFCMU
cable (1.5 m or 3.0 m)
|Z| accuracy: ± E (%)
 accuracy: ± E/100 (radian)
C accuracy:
± E (%) at Dx  0.1
2
1/2
± E(1+Dx
)
D accuracy:
± E/100 at Dx  0.1
± E(1+Dx)/100 at Dx > 0.1
G accuracy:
± E/Dx (%) at Dx  0.1
2
1/2
± E(1+Dx
)
where,
Zx: Impedance measurement value ()
Dx: Measurement value of D
E = E
' + (Z
' / |Zx| + Y
P
S
E
' = E
+ E
P
PL
POSC
Y
' = Y
+ Y
O
OL
OSC
Z
' = Z
+ Z
+ Z
S
SL
OSC
For the parameter value, see Table 4-11 to Table 4-13. Calculation example of
the measurement accuracy is also shown in Table 4-14.
Specification conditions:
Temperature: 23 C ± 5 C
Integration time: 1 PLC or 16 PLC
Agilent B1500A User's Guide, Edition 10
(%) at Dx > 0.1
/Dx (%) at Dx > 0.1
' |Zx| )  100 (%)
O
+ E
(%)
P
+ Y
(S)
O
()
S