Table 4-22
Mode
measurement range
PG mode
Fast IV mode
Table 4-23
Mode
measurement
PG mode
Fast IV mode
Table 4-24
Mode
measurement range
PG mode
Fast IV mode
Minimum rise/fall time
Current
0 to 1 V
NA
30 ns
10 mA
80 ns
1 mA
250 ns
100 A
600 ns
10 A
2 s
1 A
6 s
Settling time, at Rise time = 10 ns
Current
range
measurement
NA
10 mA
1 mA
100 A
10 A
1 A
Minimum pulse width, for pulse peak voltage = 0 to 5 V
Current
NA
10 mA
1 mA
100 A
10 A
1 A
Agilent B1500A User's Guide, Edition 10
Minimum rise/fall time
0 to 5 V
30 ns
80 ns
250 ns
600 ns
4.5 s
-
35 s
-
Settling time
Voltage
Current
measurement
150 ns
150 ns
100 ns
150 ns
250 ns
1 s
400 ns
1.2 s
10 s
6 s
80 s
Minimum pulse width
170 ns
180 ns
500 ns
1.6 s
14.5 s
115 s
Specifications
Load conditions
0 to 10 V
NA
25 pF, open
80 ns
25 pF, 1 M
250 ns
1.5 s
7 s
-
75 s
25 pF, open
Load conditions
NA
25 pF, open
25 pF, 1 k
25 pF, 10 k
25 pF, 100 k
25 pF, 1 M
25 pF, 10 M
Load conditions
25 pF, open
25 pF, 1 k
25 pF, 10 k
25 pF, 100 k
25 pF, 1 M
25 pF, 10 M