Measurement time resolution: 100 s
Standby mode
SMUs in "Standby" remain programmed to their specified output value even as
other units are reset for the next measurement.
Current offset cancel
This function subtracts the offset current from the current measurement raw
data, and returns the result as the measurement data. This function is used to
compensate the error factor (offset current) caused by the measurement path
such as the measurement cables, manipulators, or probe card.
Timestamp
The B1500A supports a time stamp function utilizing an internal quartz clock.
Resolution: 100 s
Other measurement characteristics
Measurement control:
Single, repeat, append, and stop
SMU setting capabilities:
Limited auto ranging, voltage/current compliance, power compliance,
automatic sweep abort functions, self-test, and self-calibration
Arithmetic and analysis functions
User functions
Up to 20 user-defined functions can be defined using arithmetic expressions.
Measured data and pre-defined variables can be used in the computation. The
results can be displayed on the LCD.
Arithmetic operators
+, -, *, /, ^, abs (absolute value), at (arc tangent), avg (averaging), cond
(conditional evaluation), delta, diff (differential), exp (exponent), integ
(integration), lgt (logarithm, base 10), log (logarithm, base e), mavg (moving
average), max, min, sqrt, trigonometric function, inverse trigonometric function,
and so on
Physical constants
Keyboard constants are stored in memory as follows:
Agilent B1500A User's Guide, Edition 10
Specifications
EasyEXPERT Software
4- 39