Agilent Technologies B1500A User Manual page 161

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

Selector
Agilent 16440A (B1500A-A04) SMU/pulse generator selector switches either a
SMU or PGU to the associated output port. You can expand to four channels by
adding an additional selector. The PGU port on channel 1 provides a "PGU OPEN"
function, which can disconnect the PGU by opening a semiconductor relay. Agilent
B1500A and 16445A are required to use the selector.
The following specifications data is specified at 23 C ± 5 C (73 F ± 9 F) and
50 % relative humidity.
Channel configuration: 2 channels (CH 1 and CH 2). Can add the additional 2
channels (CH 3 and CH 4) by adding another selector (selector expander).
Channel 1 (CH 1)
Channel 2 (CH 2)
Channel 3 (CH 3)
Channel 4 (CH 4)
a. PGU channels 1 and 3 have a built-in series semiconductor relay.
b. Available when a selector expander is installed.
Voltage and current range
Input port
SMU
PGU
a. This is peak-to-peak ac current. For 50  load. If DUT impedance
is lower than 50 , SPGU output voltage should be decreased for
preventing the selector from over current.
Selector Adapter
Agilent 16445A selector adapter is required to control and to supply DC power to
the selector.
Power requirement: 100 to 240 V, 50/60 Hz
Maximum volt-amps (VA): 20 VA
Agilent B1500A User's Guide, Edition 10
2 (SMU and PGU
2 (SMU and PGU)
b
2 (SMU and PGU
2 (SMU and PGU)
b
Maximum voltage (V)
200
40
Specifications
Input
a
)
a
)
Maximum current (A)
1.0
a
0.4
Hardware
Output
1
1
1
1
4- 29