Figure 3-6
Table 3-4
To Connect GNDU Adapter
The ground unit (GNDU) output is the single triaxial connector. Agilent
N1254A-100 GNDU to Kelvin Adapter enables to change it to the dual triaxial
connector. Connect the adapter to the GNDU connector on the rear panel, then
connect a Kelvin triaxial cable to extend the GNDU terminal. Maximum current is
restricted when the Kelvin triaxial cable is used. See Table 3-4.
Agilent N1254A-100 GNDU to Kelvin Adapter
Kelvin Triaxial Cable and Maximum Current
Model No. or
Option Item
16493K-001
16493K-002
Agilent B1500A User's Guide, Edition 10
Description
1.5 m Kelvin Triaxial Cable
3 m Kelvin Triaxial Cable
Installation
Installing Accessories
Maximum
Current
3.0 A
2.6 A
3- 25