Capacitance Compensation When Using Switching Matrix

Capacitance Compensation When Using
Switching Matrix
When the capacitance/conductance measurement is performed through Agilent
B2200A/B2201A/E5250A switching matrix, Agilent B1500A measures the
capacitance/ conductance of the path including a DUT (device under test), matrix
switches, extension cables and so on. So, the data measured by the B1500A is far
from the DUT's capacitance/conductance.
Agilent EasyEXPERT provides the functions used to compensate the data measured
in the environments described in "Required Conditions" on page 3-48. This section
explains how to perform compensation.
"Required Conditions"
"To Create Compensation Data File"
Corrected data by the function is not guaranteed. But typical data (supplemental
data) is as follows.
Capacitance measurement accuracy (typical): ±1 % ±0.5 pF
This typical data is for the following measurement conditions:
Measurement frequency: 1 kHz to 5 MHz
Measurement range:
Measurement terminal:
The typical data does not apply to anything extended from the 16494A/B/C cable.
The conditions described in "Required Conditions" on page 3-48 must be satisfied.
Agilent B1500A User's Guide, Edition 10
Maximum 1000 pF
At the end of Agilent 16494A/B/C cable connected to
the switch module output terminals.
Installation
3- 47