Table 3-12

Module Type and Locations

Module locations when the B1500A is shipped from the factory are shown in Table
3-12. This table shows the relative locations by the module types.
If the SPGUs are installed, the SPGUs must be installed in the slots from the slot
number 1. And if the SPGU and the WGFMU are not installed and the HPSMUs are
installed, the HPSMUs must be installed in the slots from the slot number 1. Then
the same type of modules must be installed in the contiguous slots.
For example, if the module configuration is two HPSMU, two MPSMU, and one
CMU, the B1500A will be shipped with the HPSMUs of the slots 1 to 4, the
MPSMUs of the slots 5 to 6, the CMU of the slot 7, and the blank panels of the slots
8 to 10.
Module Installation Rule
Slot Number and Location
10
:
:
:
:
:
:
:
1
0
Agilent B1500A User's Guide, Edition 10
top
B1520A Multi Frequency CMU
:
B1511A Medium Power SMU
:
:
B1517A High Resolution SMU
:
:
B1510A High Power SMU
:
B1530A WGFMU
:
:
B1525A High Voltage SPGU
:
bottom
GNDU/ADC (always installed)
Installation
About Plug-in Modules
Module Type
3- 57