Installation
Installing Accessories

Connecting the interlock terminal

The B1500A provides the Interlock connector to prevent you from receiving an
electrical shock from high voltage (more than ± 42 V). If the Interlock connector is
open, the B1500A cannot force high voltage more than ± 42 V.
When using the 16442B test fixture (B1500A-A5F), connect the Interlock
connectors together using the 16493J interlock cable.
If you use the 16495F/G/H/J connector plate, your own connector plate, or your own
test fixture, make an interlock circuit as shown in "To Make an Interlock Circuit" on
page 3-29. Then connect the B1500A Interlock connector to the external interlock
connector by using the 16493J interlock cable.