Agilent Technologies B1500A User Manual page 166

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

Table 4-19
Table 4-20
Table 4-21
Specifications
Hardware
Voltage measurement accuracy, resolution, and noise
Accuracy
b
Resolution
c
Noise
a. Independent of the range or the mode. DC constant voltage output. Appli-
cable condition: 10,000 averaging samples for 10 A range and above;
100,000 averaging samples for the 1 A range
b. Display resolution. Can vary at most 5 % based on the result of calibration
c. 0 V output, open load, no averaging. Maximum 1.5 mV
tal information
Current measurement accuracy, resolution, and noise
Accuracy
b
Resolution
Noise (Effective resolution)
a. Independent of the range or the mode. DC constant voltage output. Appli-
cable condition: 10,000 averaging samples for 10 A range and above;
100,000 averaging samples for the 1 A range
b. Display resolution. Can vary at most 5 % based on the result of calibration
c. Effective value at 0 V output, open load, and no averaging. Supplemental
information
ALWG and sampling measurement capability
Maximum number of vectors
Maximum number of sequences
Maximum number of loop counts
Length of a vector
Sampling rate
Averaging time
Hardware memory
4- 34
± (0.1 % of reading + 0.1 % of range)
680 V, -5 V to +5 V range
1.4 mV, -10 V to +10 V range
Maximum 4 mV
(-5 V to +5 V range)
rms
± (0.1 % of reading + 0.2 % of range)
0.014 % of range
Maximum 0.2 % of range
2048
512
12
1 to 10
10 ns to 10,000 s with 10 ns resolution
5 ns, or 10 ns to 1 s with 10 ns resolution
10 ns to 20 ms with 10 ns resolution
About 4 M data points/channel (typical)
Agilent B1500A User's Guide, Edition 10