Measurement mode details
Agilent B1500A supports the following measurement modes:
Staircase sweep
Multi-channel sweep (VAR1' is not supported)
Pulsed sweep
Staircase sweep with pulsed bias
IV sampling
High speed IV sampling
CV sweep
C-t sampling
C-f sweep
List sweep
Linear search (supported by FLEX command)
Binary search (supported by FLEX command)
Each SMU can be set to VAR1 (primary sweep), VAR2 (secondary sweep), VAR1'
(synchronous sweep), or CONST (constant voltage/current source).
VAR1
Primary sweep controls the staircase (DC or pulsed) voltage or current sweep.
Maximum number of VAR1 steps: N
VAR2
Subordinate linear staircase or linear pulsed sweep. After primary sweep is
completed, the VAR2 unit output value is changed.
Maximum number of VAR2 steps: N
VAR1'
Staircase or pulse sweep synchronized with the VAR1 sweep. Sweep is made
with a user specified ratio and offset values.
VAR1' = a  VAR1 + b
"a" is the user specified ratio, and "b" is the user specified offset value.
Agilent B1500A User's Guide, Edition 10
EasyEXPERT Software
= 1001
1
= 1001 (condition: 1 N
2
Specifications
N
 128128)
1
2
4- 37