Measurement mode details
Agilent B1500A supports the following measurement modes:
•
Staircase sweep
•
Multi-channel sweep (VAR1' is not supported)
•
Pulsed sweep
•
Staircase sweep with pulsed bias
•
IV sampling
•
High speed IV sampling
•
CV sweep
•
C-t sampling
•
C-f sweep
•
List sweep
•
Linear search (supported by FLEX command)
•
Binary search (supported by FLEX command)
Each SMU can be set to VAR1 (primary sweep), VAR2 (secondary sweep), VAR1'
(synchronous sweep), or CONST (constant voltage/current source).
•
VAR1
Primary sweep controls the staircase (DC or pulsed) voltage or current sweep.
Maximum number of VAR1 steps: N
•
VAR2
Subordinate linear staircase or linear pulsed sweep. After primary sweep is
completed, the VAR2 unit output value is changed.
Maximum number of VAR2 steps: N
•
VAR1'
Staircase or pulse sweep synchronized with the VAR1 sweep. Sweep is made
with a user specified ratio and offset values.
VAR1' = a VAR1 + b
"a" is the user specified ratio, and "b" is the user specified offset value.
Agilent B1500A User's Guide, Edition 10
EasyEXPERT Software
= 1001
1
= 1001 (condition: 1 N
2
Specifications
N
128128)
1
2
4- 37