Agilent Technologies B1500A User Manual page 160

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

Figure 4-2
Figure 4-3
Specifications
Hardware
Programmable timing range: 10 ns to 671.088630 ms, 10 ns resolution
SPGU supplemental information
Pulse width jitter: 0.001 % + 150 ps
Pulse period jitter: 0.001 % + 150 ps
Maximum slew rate: 1000 V/s (50  load)
Noise: 10 mVrms (at DC output)
Advanced feature:
Voltage monitor: The HV-SPGU has a voltage monitor function to measure
the voltage at the DUT terminal.
Measurement accuracy (open load): ± (0.1 % of reading + 25 mV)
Measurement resolution: 50 V
Note: Specified at 1 PLC (20 ms = (5 s sample + 5 s interval) 2000
samples)
Voltage compensation: The HV-SPGU can measure the impedance of DUT
and adjust the output voltage according to the DUT impedance.
ALWG setup table and pattern example
Pattern
P3
P4
15
10
P1
P2
P5
0
0
ALWG complex waveform example
Pattern1 x 1
4- 28
P7
P8
P6
P9
P10
500 ns
Pattern2 x 5
Pattern3 x 1
Agilent B1500A User's Guide, Edition 10
Point
Time
1
0
2
50 ns
3
70 ns
4
100 ns
5
200 ns
6
300 ns
7
320 ns
8
400 ns
9
450 ns
10
500 ns
Voltage
0.0 V
0.0 V
15.0 V
15.0 V
0.0 V
0.0 V
10.0 V
10.0 V
0.0 V
0.0 V

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents