Table 3-1
To Connect 16442B (B1500A-A5F)
Cables
Agilent 16493J Interlock Cable, 3 m or 1.5 m
Agilent 16493L GNDU Cable, 3 m or 1.5 m
Agilent 16493K Kelvin Triaxial Cable, 3 m
or 1.5 m, for the Kelvin connection
Up to 3 SMUs can be connected to the
16442B by the Kelvin connection.
Agilent 16494A Triaxial Cable, 3 m or 1.5 m,
for the non-Kelvin connection
Agilent 16493P SPGU Cable, 3 m or 1.5 m
Agilent N1300A CMU Cable
a. To extend the four-terminal pair (4TP) connection to the 16442B inputs,
connect the Hcur and Hpot cables together and the Lcur and Lpot cables
together by using the BNC T-type adapter (female-female-male, Agilent
part number 1250-2405), then connect them to the 16442B inputs.
To extend the 4TP connection to the socket module inputs, connect the
cables to the PGU1, PGU2, VSU1, VSU2, VMU1, or VMU2 connector
individually, and connect the Hcur and Hpot terminals together and the
Lcur and Lpot terminals together at the socket module inputs.
It is ok to leave open the ground wire extended from the CMU cable.
Agilent B1500A User's Guide, Edition 10
16442B
Terminals
Intlk
GNDU
SMU1
SMU2
SMU3
SMU4
SMU5
SMU6
SMU1
SMU2
SMU3
SMU4
SMU5
SMU6
PGU1
PGU2
PGU1
a
, 3 m or 1.5 m
PGU2
Installation
Installing Accessories
B1500A Terminals
Interlock
GNDU
SMU
Force
Sense
SMU
Force
Sense
SMU
Force
Sense
SMU
Force
SMU
Force
SMU
Force
SMU
Force
SMU
Force
SMU
Force
SPGU
Output 1
Output 2
CMU
Hcur/Hpot
Lcur/Lpot
3- 13