Table 2-7
Table 2-8
Introduction
Measurement Units
MPSMU Measurement Voltage Values and Resolutions
Measurement
Range
Value
0 |V| 0.55 V
0.5 V
0 |V| 2.2 V
2 V
0 |V| 5.5 V
5 V
0 |V| 22 V
20 V
0 |V| 44 V
40 V
0 |V| 100 V
100 V
MPSMU Measurement Current Values and Resolutions
Measurement
Range
Value
0 |I| 1.15 nA
1 nA
0 |I| 11.5 nA
10 nA
0 |I| 115 nA
100 nA
1 A
0 |I| 1.15 A
10 A
0 |I| 11.5 A
100 A
0 |I| 115 A
0 |I| 1.15 mA
1 mA
0 |I| 11.5 mA
10 mA
0 |I| 100 mA
100 mA
a. This column is applied to the auto ranging or the limited auto ranging.
For fixed ranging, maximum measurement value is Range column value.
2- 24
Measurement Resolutions
a
High Speed ADC
25 V
100 V
250 V
1 mV
2 mV
5 mV
Measurement Resolutions
a
High Speed ADC
50 fA
500 fA
5 pA
50 pA
500 pA
5 nA
50 nA
500 nA
5 A
Agilent B1500A User's Guide, Edition 10
High Resolution ADC
0.5 V
2 V
5 V
20 V
40 V
100 V
High Resolution ADC
10 fA
10 fA
100 fA
1 pA
10 pA
100 pA
1 nA
10 nA
100 nA