Table 4-15
Frequency range
Pulse period
Pulse width
Transition time (leading time Tr
c
and trailing time Tf)
Output relay switching time
a. Unless otherwise stated, all specifications assume a 50 termination.
b. Typical minimum. This is the supplemental information.
c. The time from 10 % to 90 % of Vamp which is the amplitude of output pulse.
d. The time it takes the open state relay to open or close.
Specifications
Hardware
Pulse range and pulse parameter
Item
Programmable range
Resolution
Minimum
Accuracy
Programmable range
Resolution
Minimum
Accuracy
Programmable range
Resolution
Minimum (typical)
Minimum
Accuracy
d
Open/close
4- 26
Value
0.1 Hz to 33 MHz
30 ns to 10 s
10 ns
100 ns (Vamp 10 V)
b
± 1 % (± 0.01 %
)
10 ns to (period -10 ns)
2.5 ns (Tr and Tf 8 s)
10 ns (Tr or Tf > 8 s)
50 ns (25 ns, typical) at Vamp 10 V
± (3 % + 2 ns)
8 ns to 400 ms (see below for the minimum
transition time)
2 ns (Tr and Tf 8 s)
8 ns (Tr or Tf > 8 s)
< 15 ns (Vamp 10 V)
20 ns (Vamp 10 V)
30 ns (Vamp 20 V)
60 ns (Vamp > 20 V)
-5 % to 5 % + 10 ns (Vamp 10 V)
-5 % to 5 % + 20 ns (Vamp 20 V)
< 100 s
Agilent B1500A User's Guide, Edition 10
a