To Connect Asu - Agilent Technologies B1500A User Manual

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

NOTE
Figure 3-1
Required parts
To install 16495K

To Connect ASU

Agilent E5288A Atto Sense and Switch Unit (ASU) will add the 1 pA range to the
high resolution SMU (HRSMU) when it is connected between the HRSMU and the
device under test (DUT). The ASU also provides the input selection function when
instruments are connected to the SMU input and the AUX input.
For the installation of the ASU and the connection to the DUT interface (probe card,
manipulators, and so on), contact your favorite prober vender. The prober vender
will have the solutions. Dimensions of the ASU are 132 mm (W)  88.5 mm (H) 
50 mm (D) excluding the connectors.
Image of ASU Installation
Agilent 16495K Connector plate
Agilent 16493M ASU cable
Agilent N1300A CMU cable or Agilent 16048D/E test leads
Agilent N1254A-108 Magnet stand
BNC cable, coaxial cable, manipulator, wire, and so on
Fix the 16495K plate to the shielding box or something that will cover the DUT
interface. See Agilent 16495 Installation Guide. The 16495K is the plate that has the
mechanism to block the light from the cable hole used to pass the cables in the
shielding box.
Agilent B1500A User's Guide, Edition 10
Installation
Installing Accessories
3- 17