Agilent Technologies B1500A User Manual page 149

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

MFCMU
MFCMU: Multi Frequency Capacitance Measurement Unit
Measurement functions
Measurement parameters: Cp-G, Cp-D, Cp-Q, Cp-Rp, Cs-Rs, Cs-D, Cs-Q,
Lp-G, Lp-D, Lp-Q, Lp-Rp, Ls-Rs, Ls-D, Ls-Q, R-X, G-B, Z-, Y-
Ranging: Auto or fixed
Measurement terminal: Four-terminal pair configuration, four BNC (female)
connectors
Cable length: 1.5 m or 3 m, automatic identification of accessories
Test signal
Frequency
Range: 1 kHz to 5 MHz
Resolution: 1 mHz (minimum)
Accuracy: ± 0.008 %
Output signal level
Range: 10 mVrms to 250 mVrms
Resolution: 1 mVrms
Accuracy:
± (10.0 % + 1 mVrms) at the measurement port of the MFCMU
± (15.0 % + 1 mVrms) at the measurement port of the MFCMU cable (1.5 m
or 3.0 m)
Output impedance: 50 , typical
Signal level monitor
Range: 10 mVrms to 250 mVrms
Accuracy (open load):
± (10.0 % of reading + 1 mVrms) at the measurement port of the MFCMU
± (15.0 % of reading + 1 mVrms) at the measurement port of the MFCMU
cable (1.5 m or 3.0 m)