Figure 2-7
Introduction
Measurement Units
HVSPGU - High Voltage SPGU
This section describes key specification of the high voltage semiconductor pulse
generator unit (SPGU) for Agilent B1500A. Maximum five modules can be
installed in one mainframe.
•
Number of output channels: 2 channels per module
Output impedance: 50
•
Output level: 0 to ± 40 V (open load), ± 20 V (50 load), 1 mV resolution
•
•
SPGU operation mode:
PG (pulse generator) or ALWG (arbitrary linear waveform generator)
•
Channel output operation mode:
Free run, duration, or pulse count (PG mode)/sequence count (ALWG mode)
•
PG output mode:
2-level pulse or 3-level pulse, for each channel
•
Pulse timing parameters: (programmable range)
Period: 20 ns to 10 s, 10 ns resolution
Width: 10 ns to period-10 ns, resolution 2.5 ns or 10 ns (transition time > 8 s)
Delay: 0 to period-20 ns, resolution 2.5 ns or 10 ns (transition time > 8 s)
Transition time (leading time or trailing time): 8 ns to 400 ms, resolution 2 ns or
8 ns (transition time > 8 ms)
•
Pulse switch, which is more durable than mechanical relays, and better suited for
frequent switching applications.
•
Automatic voltage level adjustment by specifying load impedance of DUT
SPGU Simplified Circuit Diagram
Pulse switch
50
PG 2-level
or 3-level
or
ALWG
2- 32
50
Output 1
Agilent B1500A User's Guide, Edition 10
Pulse switch
Output 2
PG 2-level
or 3-level
or
ALWG