Agilent Technologies B1500A User Manual page 170

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

Specifications
EasyEXPERT Software
CONST
A source unit can be set as a constant voltage or current source depending on the
unit.
Staircase sweep measurement mode
Forces swept voltage or current, and measures DC voltage or current. One
channel can sweep current or voltage while up to ten channels can measure
current or voltage. A second channel can be synchronized with the primary
sweep channel as an additional voltage or current sweep source.
Number of steps: 1 to 1001
Sweep mode: linear or logarithmic (log)
Sweep direction: single or double sweep
Hold time: 0 to 655.35 s, 10 ms resolution
Delay time: 0 to 65.5350 s, 100 s resolution
Pulsed sweep measurement mode
Forces pulsed swept voltage or current, and measures DC voltage or current. A
second channel can be programmed to output a staircase sweep voltage or
current synchronized with the pulsed sweep output.
Staircase sweep with pulsed bias measurement mode
Forces swept voltage or current, and measures DC voltage or current. A second
channel can be programmed to output a pulsed bias voltage or current. A third
channel can be synchronized with the primary sweep channel as an additional
voltage or current sweep source.
Sampling (time domain) measurement mode
Displays the time sampled voltage/current data (by SMU) versus time.
Sampling channels: up to 10
Sampling mode: linear or logarithmic (log)
Sampling points: 1 to 100,001/(number of channels) for linear sampling, 1 to
1 + (number of data for 11 decades) for log sampling
Sampling interval range: 100 s + 20 s  (number of channels -1) to 2 ms in
10 s resolution, or 2 ms to 65.535 s in 1 ms resolution
Hold time, bias hold time: -90 ms to -100 s in 100 s resolution, 0 to 655.35 s
in 10 ms resolution
4- 38
Agilent B1500A User's Guide, Edition 10