Agilent Technologies B1500A User Manual page 150

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

Specifications
Hardware
DC bias function
DC bias
Range: 0 to ± 25 V
Resolution: 1 mV
Accuracy: ± (0.5 % + 5.0 mV) at the measurement port of the MFCMU or
the MFCMU cable (1.5 m or 3.0 m)
Maximum DC bias current, supplemental information:
Impedance range ()
50
100
300
1
10
3
10
10
10
30
10
100
10
300
10
Output impedance: 50 , typical
DC bias monitor
Range: 0 to ± 25 V
Accuracy (open load): ± (0.2 % of reading + 10 mV) at the measurement
port of the MFCMU or the MFCMU cable (1.5 m or 3.0 m)
Sweep characteristics
Available sweep parameters: DC bias voltage, oscillator level, frequency
Sweep type: linear, log
Sweep mode: single, double
Sweep direction: up, down
Number of measurement points: maximum 1001
4- 18
Maximum DC bias current (A)
3
3
3
3
3
3
Agilent B1500A User's Guide, Edition 10
-3
10
10
-3
10
10
-3
10
10
-3
1
10
-3
1
10
-6
100
10
-6
100
10
-6
10
10
-6
10
10

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents