Specifications
Hardware
•
DC bias function
•
DC bias
Range: 0 to ± 25 V
Resolution: 1 mV
Accuracy: ± (0.5 % + 5.0 mV) at the measurement port of the MFCMU or
the MFCMU cable (1.5 m or 3.0 m)
•
Maximum DC bias current, supplemental information:
Impedance range ()
50
100
300
1
10
3
10
10
10
30
10
100
10
300
10
Output impedance: 50 , typical
•
•
DC bias monitor
Range: 0 to ± 25 V
Accuracy (open load): ± (0.2 % of reading + 10 mV) at the measurement
port of the MFCMU or the MFCMU cable (1.5 m or 3.0 m)
•
Sweep characteristics
•
Available sweep parameters: DC bias voltage, oscillator level, frequency
•
Sweep type: linear, log
•
Sweep mode: single, double
•
Sweep direction: up, down
•
Number of measurement points: maximum 1001
4- 18
Maximum DC bias current (A)
3
3
3
3
3
3
Agilent B1500A User's Guide, Edition 10
-3
10
10
-3
10
10
-3
10
10
-3
1
10
-3
1
10
-6
100
10
-6
100
10
-6
10
10
-6
10
10