Table 4-18
Voltage force accuracy, resolution, and timing
V force (Fast IV mode)
V force (PG mode)
Accuracy
b
Resolution
Overshoot/undershoot
Noise
Rise time T
(10 to 90 %)/
rise
(90 to 10 %)
Fall time T
fall
Pulse period
Pulse width
a. Independent of the range or the mode. DC constant voltage output. Load
impedance must be  1 M (1 A range) or  200 k (all other current
ranges) for Fast IV mode, or  1 M for PG mode.
b. Can vary at most 5 % based on the result of calibration
c. PG mode, 50  load, T
with 3 m cable, or >56 ns with 5 m cable
d. Theoretical value for the observed time 100 ns to 1 ms, supplemental
information
e. PG mode, 50  load, T
f. PG mode, 50  load, pulse period  100 ns
g. PG mode, 50  load, pulse width  50 ns
Agilent B1500A User's Guide, Edition 10
-5 V to +5 V, -10 V to 0 V, 0 V to +10 V
-5 V to +5 V, open load
-2.5 V to +2.5 V, 50  load
± 0.1 % of setting ± 0.1 % of range
96 V (-3 V to +3 V)
160 V (all ranges except for -3 V to +3 V)
c
± (5 % + 20 mV)
Maximum 0.1 mV
rms
Accuracy: -5 % to +(5 % + 10 ns) of setting
Minimum: 24 ns, PG mode and 50  load
Timing accuracy: ± 1 % of setting
Minimum: 100 ns, PG mode and 50  load
Accuracy: ± (3 % + 2 ns)
Minimum: 50 ns, PG mode and 50  load
and T
>16 ns with the 1.5 m cable, >32 ns
rise
fall
 24 ns
and T
rise
fall
Specifications
Hardware
a
d
f
g
e
4- 33