Table 3-9
Compensation Coefficients and Modifications
Compensation
Agilent B1500A User's Guide, Edition 10
Capacitance Compensation When Using Switching Matrix
For E5250A:
# E5250A
C Compensation coefficient data table
#
#
CAUTION : Do not add or delete "REVISION" line and "DATAxx" line.
#
Change the value for R,L,C of DATA05,06,07 or 08.
#
REVISION
A.03.00
#
R [ohm]
DATA00
74.65E-3
DATA01
75.41E-3
DATA02
231.41E-3
DATA03
177.56E-3
DATA04
100.70E-3
DATA05
100.70E-3
DATA06
100.70E-3
DATA07
114.00E-3
DATA08
114.00E-3
DATA09
0.00E-3
# END of Data
The lines DATA05 to DATA08 should be modified for each measurement
environment. See Table 3-9. Do not modify the other lines.
coefficients
C2H
For the connector plate, change the R, L, C values in the lines.
(DATA05)
The value must be changed to the R, L, C values of the
C2H (DATA05) path and the C2L (DATA06) path (triaxial
C2L
cable with connector plate) shown in Figure 3-18.
(DATA06)
For Agilent B2220A probe card interface, do not modify the
lines.
C3H
Change the R, L, C values in the lines. The value must be
(DATA07)
changed to the R, L, C values of the C3H (DATA07) path and
the C3L (DATA08) path shown in Figure 3-18.
C3L
(DATA08)
For the connector plate, the coaxial cable with positioner will
be the appropriate path.
For Agilent B2220A probe card interface, the probe card will
be the appropriate path.
L [H]
C [F]
140.00E-9
58.44E-12
90.00E-9
67.13E-12
450.00E-9
178.85E-12
390.00E-9
135.45E-12
400.00E-9
80.00E-12
400.00E-9
80.00E-12
400.00E-9
80.00E-12
544.00E-9
130.00E-12
544.00E-9
130.00E-12
0.00E-9
1.20E-12
Modifications of data file
Installation
# Frame Path 1
# Frame Path 2
# Card Path High
# Card Path Low
# Triax Cable
[/m]
# User Triax Cbl H [/m]
# User Triax Cbl L [/m]
# User Coax Cbl H
[/m]
# User Coax Cbl L
[/m]
# Stray Capacitance