Agilent Technologies B1500A User Manual page 56

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

WARNING
CAUTION
Introduction
Rear View
There are potentially hazardous voltages (200 V for HPSMU, and 100 V for
MPSMU/HRSMU/ASU) present at the Force, Sense, and Guard terminals of
Agilent B1500A. To prevent electrical shock, the following safety precautions
must be observed during the use of the B1500A.
a. Use a three-conductor AC power cable to connect the cabinet (if used)
and the B1500A to an electrical ground (safety ground).
b. If you do not use the 16442B test fixture (B1500A-A5F), you must install
and connect an interlock circuit that opens the Interlock terminal when
the shielding box access door is opened.
c. Confirm periodically that the interlock function works normally.
d. Before touching the connections on the Force, Guard, and Sense
terminals, turn the B1500A off and discharge any capacitors. If you do
not turn the B1500A off, complete all of the following items, regardless
of the B1500A settings.
Press the
Stop
Confirm that the High Voltage indicator is not lit.
Open the shielding box access door (open the Interlock terminal).
Discharge any capacitors connected to an SMU.
e. Warn persons working around the B1500A about dangerous conditions.
E. CMU module
The multi frequency capacitance measurement unit (MFCMU) has four coaxial
connectors, Lcur, Lpot, Hpot, and Hcur, for the four-terminal pair connection.
The Hcur and Hpot have to be connected together at a terminal of the device
under test (DUT). Also the Lcur and Lpot have to be connected together at the
other terminal of the DUT. Then, the guard lines of four-terminal pair must be
connected together.
Do not apply voltage more than ±25 V to the MFCMU input terminals. Failure to
heed this caution may result in damage to the module.
2- 16
key to set the module output off.
Agilent B1500A User's Guide, Edition 10