Repairing A 2.4 Mm Rf Connector - Agilent Technologies 8511B Operating And Service Manual

Frequency converter test set
Table of Contents

Advertisement

(4) Repairing a
2.4 mm RF Connector
Figure 7-2
Replacement Procedures
Refer to
Figure 7-2
and the following text to repair a 2.4 mm connector.
Refer to the "Replaceable Parts" section of this manual for replacement part
numbers.
Disassembly
Exploded Diagram of a 2.4 mm Connector
1. Remove any attached cables from the connector to be replaced.
2. From inside the test set, use a 9/16-inch nut driver to loosen the nut and
remove the connector.
Assembly
3. Assemble the bulkhead connector, lock washer and nut.
4. Using the 9/16-inch nut-driver, tighten the nut to 506 N-cm (45 in-lb).
5. Attach the assembly to the bulkhead connector. Torque to 281 N-cm
(25 in-lb).
6. Clean the mating surfaces with alcohol and lint-free swabs.
7. Check the pin depth of the gold nose connector. The pin depth
specification is 0.0000 to +0.0030 inch.
8. Reconnect the cables disconnected in step 1.
7-6 Agilent 8511B Test Set Operating and Service Manual