There Appears To Be A Plate Present - Agilent Technologies AssayMAP Bravo Platform Manual

Error recovery guide
Table of Contents

Advertisement

2 Common startup error messages

There appears to be a plate present

There appears to be a plate present
Startup error message
Resolving the error
Cause
The first time that the AssayMAP Bravo device
is initialized after turning on the instrument, this
error occurs as a prompt to make sure the
gripper is not holding a plate.
4
The There appears to be a plate present error message appears during initialization of
the AssayMAP Bravo Platform.
Recommended action
1
Make sure that no labware is being held by the
AssayMAP Bravo gripper.
Note: If the gripper is holding labware, open Bravo
Diagnostics, and use the controls in the Gripper tab
to open the gripper and remove the labware.
2
Click Ignore and Continue, leaving device in current state
to continue the homing process.
AssayMAP Bravo Platform Error Recovery Guide