NOTE
Agilent B1500A has the following front panel user interface.
Stop
Stops the present measurement or source output immediately.
Rotary knob
Rotating the knob moves the marker on the graph window, or
increases/decreases/changes the value in the active entry field.
Pressing the knob sets or enters the value.
Softkeys
Seven softkeys are available. Used to select one for the entry field specified or
the dialog box. They are also used to recall the test definitions (pre-defined test
setups).
Touch Panel Off
Enables or disables the touch screen operation.
Standby switch
Turns the B1500A on. Pressing the button in the ON state makes the B1500A in
the standby state.
LCD Off
Enables or disables the LCD panel. LED lights when the LCD is disabled.
Auto, -, +, Menu
Four keys are available for the display adjustment. Press Auto to perform
adjustment automatically. Use -, +, and Menu to adjust manually.
Agilent B1500A also requires Agilent EasyEXPERT software as its graphical user
interface. You can operate Agilent EasyEXPERT by using the touch panel. Use your
fingers, stylus pen, and so on for the touch panel operation. The USB keyboard and
the USB mouse are also available for operating the EasyEXPERT.
To remove USB devices from the B1500A, use "Safely Remove Hardware" on
Windows taskbar. If it is not used, the B1500A may cause the internal USB
communication error. If the error occurs, turn the B1500A off and disconnect the
power cable from the B1500A. Leave it about 30 seconds before rebooting the
B1500A, and connect the power cable again, and then turn the B1500A on.
Agilent B1500A User's Guide, Edition 10
Getting Started
To Turn On/Off B1500A
1-5