Specifications
Hardware
MPSMU/HRSMU
MPSMU: Medium Power Source/Monitor Unit
HRSMU: High Resolution Source/Monitor Unit
Voltage range, resolution, and accuracy
See Table 4-1 for high resolution ADC and Table 4-3 for high speed ADC.
Current range, resolution, and accuracy
See Table 4-2 for high resolution ADC and Table 4-4 for high speed ADC.
Specified measurement resolution is limited by fundamental noise limits.
Minimum displayed resolution is 1 aA at 1 pA range by 6 digits. Measurements
made in the lower ranges can be greatly impacted by vibrations and shocks.
These specifications assume an environment free of these factors.
Power consumption
Voltage source mode:
Voltage range (V)
0.5
2
5
20
40
100
a. Ic is the current compliance setting.
Current source mode:
Voltage compliance
Vc  20
20 < Vc  40
40 < Vc  100
a. Vc is the voltage compliance setting.
b. Io is the output current.
4- 6
Power consumption
Ic  20
Ic  20
Ic  20
Ic  20
Ic  40
Ic  100
a
(V)
Power consumption
Io  20
Io  40
Io  100