Figure 2-8
WGFMU - Waveform Generator/Fast Measurement
Unit
This section describes typical specification of the waveform generator/fast
measurement unit (WGFMU) and the remote-sense and switch unit (RSU) for
Agilent B1500A. Maximum five modules can be installed in one mainframe.
Number of channels: 2 channels per module
Function:
Voltage output and current or voltage sampling measurement with minimum
sampling interval of 5 ns
Current measurement range: 1 A, 10 A, 100 A, 1 mA, or 10 mA
Operation mode: PG mode, Fast IV mode, or DC mode
PG mode:
ALWG (arbitrary linear waveform generator) voltage output and voltage
measurement (VFVM).
Output level: -5 V to 5 V for open load or -2.5 V to 2.5 V for 50  load
Minimum pulse width: 100 ns
Fast IV mode:
ALWG voltage output and current or voltage measurement (VFIM or VFVM).
Output level: 0 to -10 V, 0 to 10 V, or -5 V to 5 V
Minimum pulse width: 300 ns
DC mode:
DC voltage output and current or voltage measurement (VFIM or VFVM).
Output level: 0 to -10 V, 0 to 10 V, or -5 V to 5 V
WGFMU Simplified Circuit Diagram
PG mode
50
ALWG
A
Fast IV mode
Agilent B1500A User's Guide, Edition 10
V
Ch1
RSU output
ALWG
V
Measurement Units
PG mode
50
V
A
V
Fast IV mode
Introduction
Ch2
RSU output
2- 33