Error Information
Refractive Index Detector Specific Error Messages
Lamp Voltage too Low
Error ID: 2693
Probable cause
1
2
Lamp Voltage too High
Error ID: 2690
Probable cause
1
2
3
Lamp Current too High
Error ID: 2691
Probable cause
1
2
Agilent InfinityLab LC Series RID User Manual
Defective main board.
Defective lamp or optics.
Contaminated flow cell.
Defective main board.
Defective lamp or optics.
Defective main board.
Defective lamp or optics.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Flush the flow cell.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.