Lamp Voltage Too Low; Lamp Voltage Too High; Lamp Current Too High - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
Refractive Index Detector Specific Error Messages

Lamp Voltage too Low

Error ID: 2693
Probable cause
1
2

Lamp Voltage too High

Error ID: 2690
Probable cause
1
2
3

Lamp Current too High

Error ID: 2691
Probable cause
1
2
Agilent InfinityLab LC Series RID User Manual
Defective main board.
Defective lamp or optics.
Contaminated flow cell.
Defective main board.
Defective lamp or optics.
Defective main board.
Defective lamp or optics.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Flush the flow cell.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
99

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents