Siemens MASS 2100 Operating Instructions Manual page 72

Sitrans f coriolis flowmeters
Hide thumbs Also See for MASS 2100:
Table of Contents

Advertisement

Connecting
5.2 Connecting MASS 2100/FC300
6. Connect wires to terminals, wall mount transmitter.
7. Tighten cable gland.
Note
For Ex versions active or passive current output is preselected at ordering and cannot be
changed.
Non-Ex versions can be connected as either active or passive.
Note
Load
Signal output: < 500 Ω at 14 to 24 VDC (active), 14 to 30 VDC (passive)
Relay output: 30 VAC/VDC, 100 mA
Passive signal input: 15 to 30 VDC, 2 to 15 mA
Table 5-1
Load [Ω] Voltage (active
100
200
500
1000
2000
5000
10000
20000
50000
100000
72
Active current output
5
4
HW polling address
Voltage (active non
Ex) [V]
Ex) [V]
3
3
5.9
5.9
11.4
14.9
14.8
19.2
17.4
20.1
19.4
20.4
20.3
21
20
20.4
20.4
20.6
20.6
20.7
Ca+
C
24 V DC Voltage supply (passive) [V] Ex and
Non Ex
Measurered
17.7
19.6
21.0
21.6
21.9
22.2
22.3
22.4
22.5
22.6
Operating Instructions, 06/2017, A5E39789040-AA
Passive current output
6
Cp-
5
C
MASS 2100 & FC300 (FCT030)