HMI menu ID
Parameter
2.4.3.6.11
Fail-safe minimum time
2.4.3.7
Pulse output
2.4.3.7.1
Redundancy mode
2.4.3.7.2
Process value
2.4.3.7.3
Direction
2.4.3.7.4
Pulse units
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
Description
Minimum time the output stays
in fail-safe mode. Short events
will be prolonged.
Set the redundancy mode. The
pulses on channel 3 will follow
pulses on channel 2 with 90 or
180 deg. shift.
Select process value for current
output.
Select measured flow direction.
The flow meter will only output
current when flow value is meas‐
ured in selected direction.
Select units for pulse output.
Units are not applied at commu‐
nication interface.
Mass flow
B.5 Inputs and outputs
Default value
Value range
[unit]
0
● 0 to 100 s
Off
● Off
● 90º
● 180 º
Mass flow
● Mass flow
● Volume flow
● Standard
volume flow
● Fraction flow
A
● Fraction flow
B
Positive
● Positive
● Negative
● Bidirectional
kg
● g (grams)
● kg (kilograms)
● t (1 t = 1000
Kg)
● STon (1 STon
= 2000 lb)
● T (1 T = 2240
lb)
● oz (ounces)
● lb (pounds)
● Custom
Default settings
Access
level
User
User
User
User
User
315