Diagnosing and troubleshooting
10.2 Fault codes and corrective actions
ID/
Diagnostic
Icons
110
Medium temperature
111
Medium temperature
112
Fraction A %
113
Fraction A %
114
Fraction A %
115
Fraction A %
116
Fraction B %
117
Fraction B %
118
Fraction B %
196
Action
Value below warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter 'Lower warning limit'
Value below alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter 'Lower alarm limit'
Value above alarm limit. Check process condi‐
tions. Align Upper alarm limit to normal process
conditions.
Value above warning limit. Check process con‐
ditions. Align Upper warning limit to normal
process conditions
Value below warning limit. Check process con‐
ditions. Align Lower warning limit to normal
process conditions.
Value below alarm limit. Check process condi‐
tions. Align Lower alarm limit to normal process
conditions.
Value above alarm limit. Check process condi‐
tions. Align Upper alarm limit to normal process
conditions.
Value above warning limit. Check process con‐
ditions. Align Upper warning limit to normal
process conditions
Value below warning limit. Check process con‐
ditions. Align Lower warning limit to normal
process conditions.