Siemens MASS 2100 Operating Instructions Manual page 225

Sitrans f coriolis flowmeters
Hide thumbs Also See for MASS 2100:
Table of Contents

Advertisement

Positive flow across zero with positive scaling
Current output setting
● Process value = Massflow
● Direction = Bidirectional
● Current Mode = 4-20 mA NAMUR
● Upper Scaling = 400 kg/h
● Lower Scaling = -100 kg/h
● Fail Safe Mode = Maximum current
● Low-Flow Cut-Off = 25 kg/h
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
mA
-100 kg/h
Lower scaling
Low-flow cut-off
Upper scaling
Maximum measurement value
Upper range
Lower range
Minimum measurement value
Lower alarm value
Measurement range
Kg/h
400 kg/h
Technical data
11.5 Outputs
225

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fc300Fct030

Table of Contents