Default settings
B.5 Inputs and outputs
HMI menu ID
Parameter
2.4.3.7.4
Pulse units
2.4.3.7.4
Pulse units
2.4.3.7.5
Amount
2.4.3.7.6
Pulses per amount
316
Description
Select units for pulse output.
Units are not applied at commu‐
nication interface.
Volume flow
Select units for pulse output.
Units are not applied at commu‐
nication interface.
Standard volume flow
Define the amount per pulse, or
multiple pulses if you configure
more than 1 pulse per amount
Define number of pulses which
will be output when configured
amount is reached
Default value
Value range
[unit]
l (liters)
● l (liters)
● hl (hectoliters)
● m
(cubic
3
meters)
● gal (US
gallons)
● gal (UK)
(imperial
gallons)
● bbl (US liquid
barrels = 31.5
US gallons)
● bbl-beer (beer
barrels = 31
US gallons)
● bbl-beer (beer
barrels = 42
US gallons)
● In
(cubic
3
inches)
● ft
(cubic feet)
3
● yd
3
yards)
● bu (bushels)
● Custom
Sl (standard lit‐
● Nl (normal
ers)
liters)
● Nm
3
cubic meters)
● SI (standard
liters)
● Sft
3
cubic feet)
● Sm
3
cubic meters
● Custom
1
1
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
Access
level
User
(cubic
(normal
(standard
(standard
User
User