Default settings
B.5 Inputs and outputs
HMI menu ID
Parameter
2.4.5.8.16
NAMUR status signal
2.4.5.8.17
Polarity
2.4.5.8.18
On delay
2.4.5.8.19
Off delay
2.4.5.9
Digital input
2.4.5.9.1
Input function
2.4.5.9.2
CH1 forced current val‐
ue
2.4.5.9.3
CH2 forced value
362
Description
Select one or more NAMUR sta‐
tus signals. The output is active
when one of the selected signals
occurs.
Set the level of an active signal. Active high level ● Active high
Set the time delay between
alarm condition and activating
the output. Allows to ignore tran‐
sient fault conditions
Set the time delay between re‐
setting alarm condition and de-
activating the output. Allows to
ignore transient fault conditions
Select the signal input function‐
ality.
Enter the value for the output
signal when forcing is activated.
Enter the value for the output
signal when forcing is activated.
Current
Default value
Value range
[unit]
● Failure
● Function
check
● Out of
specification
● Maintenance
required
● None
level
● Active low
level
0 s
● 0 to 100 s
0 s
● 0 to 100 s
● Start dosing
● Stop dosing
● Reset totalizer
1
● Reset totalizer
2
● Reset totalizer
3
● Reset all
totalizers
● Zero point
adjustment
● Resume
dosing
● Force outputs
● Freeze
process
values
4.00 mA
3.5 to 25 mA
4.00 mA
3.5 to 25 mA
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
Access
level
User
User
User
User
User
User
User