Technical data
11.5 Outputs
Positive flow with negative scaling
Current output setting
● Process value = Mass flow
● Direction = Symmetric
● Current Mode = 4-20 mA (maximum 25 mA)
● Upper Scaling = 100 kg/h
● Lower Scaling = 400 kg/h
● Fail Safe Mode = Maximum current
● Low-Flow Cut-Off = 25 kg/h
224
mA
100 kg/h
Low-flow cut-off
Upper scaling
Maximum output current
Upper alarm current
Upper range
Lower scaling
Lower range
Minimum output current
Measurement range
kg /h
400 kg/h
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA