Default settings
B.5 Inputs and outputs
HMI menu ID
Parameter
2.4.3.6.2
Process value
2.4.3.6.2
Direction
2.4.3.6.4
Frequency value high
2.4.3.6.5
Frequency value low
2.4.3.6.6
Upper range value
2.4.3.6.7
Lower range value
2.4.3.6.8
Damping value
2.4.3.6.9
Fail-safe behavior
2.4.3.6.10
Fail-safe value
314
Description
Select measured flow direction.
The flow meter will only output
current when flow value is meas‐
ured in selected direction.
Assign maximum frequency val‐
ue
Assign minimum frequency val‐
ue
Assign process value to the
nominal upper range value
Assign process value to the
nominal lower range value
Set filter time constant. Defines
the damping of the current out‐
put signal. 0.0 means filter is off
Select current output reaction in
case of a fault.
Define fail-safe current output
value if 'Fail-safe behaviour' pa‐
rameter is set to 'Fail-safe value'.
Default value
Value range
[unit]
Mass flow
● Mass flow
● Volume flow
● Standard
volume flow
● Density
● Standard
density
● Medium
temperature
● Fraction A
● Fraction B
● Fraction A %
● Fraction B %
● Control valve
Positive
● Positive
● Negative
● Bidirectional
● Symmetric
10000
● 0 to 12500 Hz
0
● 0 to 12500 Hz
1)
1)
1)
1)
0 s
● 0 to 100 s
Last reliable val‐
● Minimum
ue
frequency
● Maximum
frequency
● Last reliable
value
● Current value
● Fail-safe
value
0 Hz
● 0 to 12500 Hz
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
Access
level
User
User
User
User
User
User
User
User
User