Arterial Pressure Transducer Calibr - Gambro AK 95 S Service Manual

Dialysis machines
Table of Contents

Advertisement

Calibrations
Arterial Pressure Transducer Calibration
1.
Connect the Gambro reference instrument
(together with a calibration tube set) to
the arterial pressure transducer on the
front of the AK 95 S, see fig. below.
2.
Zeroset the reference instrument, wait for
stable reading.
3.
Start the AK 95 S, enter the Service
menu and select Calibrations.
4.
Select Transducers, External and
ARTERIAL PRESSURE TRANSDUCER
from the selection diagram.
5.
Press
to confirm the selected trans-
ducer.
6.
Apply zero pressure by opening the three
clamps A, B and C on the calibration tube
set. Set the "REF SET" value on the
AK 95 to 0 mmHg.
7.
Press
to enter the setting.
START
8.
If the lit
UF
STOP
the machine stores the offset for the
transducer in the EEPROM and updates
the checksum.
For the sensitivity calibration
(the slope):
Close clamp A. Apply a pressure of approx.
-500 mmHg, shown on the reference in-
strument. Close clamp B.
ART PR
LOW
1.2
-550
9.
Set the "REF SET" value on the AK 95 S
to the reading on the reference instrument
and press
10.
Open clamp A and B, then close clamp A
again. Apply a pressure of approx. +500
mmHg, shown on the reference instru-
ment. Close clamp B.
6:6
key is pressed at this time
REF SET
HIGH
-500
550
twice.
ART PR
LOW
1.2
-550
11.
Set the "REF SET" value on the AK 95 S
to the reading on the reference instrument
and press
.
START
12.
Press
to load the calibration values
UF
STOP
into the EEPROM.
13.
When the EEPROM storage is ready,
press
to exit.
If a calibration error occurs:
Try to do the calibration again.
If the error occurs a second time:
Change the transducer and the transducer
board.
Report to your Gambro service representa-
tive the error message on the error dis-
play.
B
A
REF SET
HIGH
+500
550
C
AK 95 / AK 95 S Service Manual
HCEN 9138 Rev. 10.2005