Degassing Pressure Transducer Calibration
1.
Disconnect tube no. 50 at the Expansion
chamber. Connect the Gambro reference
instrument to tube no. 50 via the calibra-
tion tube set. Use the angled-silicone
connector from the calibration kit.
2.
Disconnect tube no. 11 at the degassing
chamber or at the outlet of the flow pump
and plug the silicone connector. Use a plug
from the calibration kit.
4.
Zeroset the reference instrument, wait for
stable reading.
5.
Start the AK 95 S, enter the Service
menu and select Calibrations.
6.
Select Transducer, External and
DEGASSING PRESSURE TRANSDUCER
from the selection diagram.
7.
Press
to confirm the selected trans-
ducer.
8.
Apply zero pressure by opening the two
clamps A and B on the calibration tube
set. Set the "REF SET" value on the
AK 95 S to 0 mmHg.
9.
Press
to enter the setting.
10.
Close clamp A. Apply a pressure of approx.
-500 mmHg, shown on the reference
instrument. Close clamp B.
11.
Set the "REF SET" value on the AK 95 S
to the reading o the reference instrument
and press
12.
Open clamp A and B, then close clamp A
again.
Apply a pressure of approx. +500 mmHg,
shown on the reference instrument. Close
clamp B.
AK 95 / AK 95 S Service Manual
HCEN 9138 Rev. 10.2005
twice .
13.
Set the "REF SET" value on the AK 95 S to
the reading on the reference instrument
and press
.
START
14.
Press
to load the calibration values
UF
STOP
into the EEPROM.
15.
When the EEPROM storage is ready,
press
to exit.
If a calibration error occurs:
Try to do the calibration again.
If the error occurs a second time:
Change the transducer and the transducer
board.
Report to your Gambro service representa-
tive the error message on the error dis-
play.