Dialysis machines
Table of Contents

Advertisement

Service Technicians Guide
Technical error handling
For more information about the error message
refer to the Error Code List for AK 95 S,
HCEN 9091.
Note! It is important that the program version
on the Error Code List corresponds with the
AK 95 S PROM.
If a technical error appears, the AK 95 S places
an error code in a memory, called the error code
buffer. The buffer's content is kept after the
machine is switched off, allowing the technician
to view it at a later time.
When an error is generated, the machine
alarms, and the buffer can be accessed by press-
ing the Technical Error button.
The main groups can be interpreted as follows:
C CB ... are generated by Blood functions
C CC ... are generated by Common functions
C CF ... are generated by Fluid functions
C CO ... are generated by OpCom
P PTRA ... are generated by Ptransducer
functions
1)
OpCom=Operator's Communication
2)
PTransducer= Protective system transducers.
All different error codes are available in the
Error Code List.
4:6
1)
functions
2)
The position of the error in the buffer.
Total numbers of errors in the buffer.
Buf:
04
(07)
C COF
088 031
The error code.
27 Oct 05 6:43
Time stamp: When
did it happen.