Gambro AK 95 S Service Manual page 106

Dialysis machines
Table of Contents

Advertisement

Test equipment
Test equipment needed to perform maintenance.
Test equipment:
Conductivity measurement
Pressure measurement
Temperature measurement:
Digital Voltmeter
Measuring glass
Blood leak Calibration
Cover
Gauge pin kit for blood pump rollers
(including three different pins)
Gauge pin kit for blood pump rollers
(to be ordered separately)
Calibration tube kit
Drip chamber
Tube for calibration of
priming detector
Safety tester
*BPM test equipment
*BPM Cuff
*pH buffers
Magnet for blood pump cover
* If installed
AK 95 / AK 95 S HCEN12189 Rev.
10. 2006
Tolerance:
± 0.1 mS/cm
± 2mmHg to ± 200mmHg
± 1% beyond ± 200mmHg
If BPM is installed:
± 0.8mmHg
± 0.2ºC
± 0.5%
± 5 ml
N/A
T = 1.1 (Stop pin = 1.1*2*0.7 = 1.5mm)
T = 1.6 (Stop pin = 1.6*2*0.7 = 2.2mm)
T = 2.0 (Stop pin = 2.0*2*0.7 = 2.8mm)
T = wall thickness of the blood tubing
T = 1.1 (Stop pin = 1.1*2*0.7 = 1.5mm)
T = 1.6 (Stop pin = 1.6*2*0.7 = 2.2mm)
T = 2.0 (Stop pin = 2.0*2*0.7 = 2.8mm)
T = wall thickness of the blood tubing
N/A
N/A
N/A
According to IEC 60601-1
N/A
N/A
± 0.07 pH units
Manufacturer:
Gambro Order
number:
MESA, IBP
MESA, IBP, Druck
MESA, IBP
Fluke
Gambro
K40169001
Gambro
K40158001
Gambro
K40158002
K40158003
K40158004
Gambro
K13983002
Rigel, Biotec,
Metron
Gambro
K22151001
Gambro
100900141
pH 7 – 100810001
pH 9 – 100810002
Gambro
K19049001
Maintenance Manual
5:3