Service Technicians Guide
Preset mode
In the Preset mode the set up of AK 95 S is
confirmed. Following choices are set in preset
mode:
• Start up values
• Configuration
• Set values and limits
• Override times
• Auto priming parameters
• Rinse Back
• Disinfection parameters
• Single needle paramers
• Acetate concentrates
• Bicarbonate concentrates
• BPM settings
• Profiling
Enter preset mode as follows:
MAINTENANCE SUPPORT SELECTION
Logging
Service/Preset
Select Service/Preset
MAINTENANCE SELECTION
Fch
Select Preset and following dispalys shows up:
INITIALIZING PRESET MODE
SELECT PRESET TYPE
Manual
Ext Default
Select "Manual" when AK 95 S is preseted from
the operatorspanel and select "Ext" when the
AK 95 S is connected to a PC and preseted via
the Gambro software GXP.
It is also possible to preset a AK 95 S machine
from another AK 95 S machine with a Monitor to
Monitor preset cable, K2 2753 001.
4:8
Preset
Service
AK 95 / AK 95 S Service Manual
HCEN 9136 Rev. 10.2005